Instruments that say not just what is in a sample but how much, often down to a few parts per million. The colorimeter and spectrophotometer measure how much light a coloured solution swallows; the flame photometer measures sodium and potassium by the colour they give a flame; the polarograph reads metals from the current at a dropping mercury drop; the mass spectrometer weighs single atoms. How to build each, how liquid chromatography works in outline, and how to calibrate any of them with standards.

Difficulty
●●●●●
Time
A weekend for a filter colorimeter; a month for a flame photometer or a spectrophotometer; a mass spectrometer is a project of a year for a skilled team
Steps from zero
57 prerequisites →

You need first

ChromatographyThe SpectroscopePhotocells and Sensors

This unlocks

—

Raw materials

saltquartz
The full path from nothing — 57 steps before this one
  1. Stone Tools
  2. Fire
  3. Cordage and Rope
  4. Finding and Preparing Clay
  5. Finding and Identifying Ores
  6. Sulfur
  7. Simple Machines
  8. Hunting and Trapping
  9. Basic Woodworking
  10. Quarrying and Cutting Stone
  11. Pottery
  12. The Kiln
  13. Charcoal
  14. Hides and Leather
  15. Mining and Tunnelling
  16. Lime
  17. Bricks
  18. Salt
  19. Mercury
  20. Fermentation
  21. Lye and Potash
  22. The Wheel and Axle
  23. Glass
  24. Firebrick and Refractories
  25. Bellows
  26. Distillation
  27. Saltpeter
  28. Acids, Bases and Testing
  29. The Lathe
  30. Gears
  31. Coal and Coke
  32. Copper Smelting
  33. Iron (the Bloomery)
  34. Coal Tar Chemicals
  35. Laboratory Glassware
  36. Sulfuric Acid
  37. Pumps
  38. The Waterwheel
  39. Lenses
  40. Blacksmithing
  41. Organic Synthesis and Fine Chemicals
  42. Hydrochloric Acid
  43. Identifying Substances
  44. The Battery
  45. The Vacuum Pump
  46. Magnets
  47. The Spectroscope
  48. Chromatography
  49. Selenium and Tellurium
  50. The Light Bulb
  51. Copper Wire
  52. The Generator
  53. The Vacuum Tube
  54. The Electric Arc Furnace
  55. Pure Silicon
  56. Semiconductors and the Transistor
  57. Photocells and Sensors
  58. Analytical Instruments

Do them roughly in this order. See it on the tech tree →

Why it matters

Identifying Substances tells you what is in a sample. Most real questions ask how much. Is there 0.3 mg of iron per litre of drinking water, or 3? Is a patient's blood sodium 140 or 120 millimoles a litre? Does this batch of steel hold 0.4 % manganese or 1.4 %? Has the fertiliser enough potassium?

Weighing and titration (Acids, Bases and Testing) answer these when there is plenty of the substance. Instruments answer them for traces, quickly, dozens of samples a day, in waterworks (Water Treatment for a Town), hospitals (The Medical Laboratory), steelworks and fertiliser plants (Chemical Fertilisers).

Every instrument here turns the amount of a substance into an electric current, measures it with a meter (Measuring Electricity), and compares the reading with standards of known strength. The last step matters most.

How it works

Coloured solutions swallow light. A copper solution looks blue because it swallows orange and red. The more copper, or the longer the path through it, the more is swallowed.

The rule, Beer's law, is simple if you use the right measure. Call the light going in I₀ and the light coming out I. The transmittance T = I / I₀. The absorbance A = log₁₀(I₀ / I). Then:

A = ε × c × l

where c is the concentration, l the path length through the liquid, and ε a number belonging to the substance and the colour of light. In plain words: absorbance is proportional to concentration. Double the iron, double the absorbance.

Light let through (T) Absorbance (A)
100 % 0
50 % 0.30
10 % 1.0
1 % 2.0

Readings are most accurate between about A = 0.2 and 0.8 (T from 65 % down to 15 %). Dilute or concentrate the sample, or change the path length, to land there.

Use the colour the solution swallows. Measure with light of the colour the solution absorbs most, which is the opposite of the colour it looks.

Solution looks Measure with Wavelength
Yellow Blue or violet 420–450 nm
Orange or red Blue-green 470–500 nm
Purple (permanganate) Green 520–540 nm
Blue (copper) Orange-red 600–650 nm

Atoms in a flame give their own colours. Sodium in a flame shines yellow at 589 nm, potassium deep red at 766 nm, lithium crimson at 671 nm (The Spectroscope). The brightness is, over a useful range, proportional to the amount sprayed in. That is the flame photometer.

Standards and calibration

Lamps age, photocells tire, flames flicker: no instrument gives a true answer by itself. Each reading is compared, the same day, with standard solutions of known strength.

Making standards

  1. Dry a pure salt at 105–110 °C for two hours and cool it in a desiccator (Burners, Stands and Laboratory Equipment).
  2. Weigh it on your best balance (Balances, Steelyards and Weighing Machines) and dissolve it in distilled water (Distillation) in a volumetric flask, a flask with one etched mark on its narrow neck that holds exactly 1 litre or 100 ml when filled to the mark at 20 °C (Laboratory Glassware).
To make 1,000 mg per litre of Dissolve in 1 litre
Sodium 2.542 g sodium chloride
Potassium 1.907 g potassium chloride
Iron 7.022 g ferrous ammonium sulfate, plus 5 ml strong sulfuric acid
Copper 3.929 g copper sulfate crystals (CuSO₄·5H₂O)
  1. Dilute this stock to make working standards. With a pipette, take 10 ml into a 1-litre flask and fill to the mark: 10 mg/L. From that, take 10, 20, 50 ml into 100 ml flasks: 1, 2 and 5 mg/L. Write the date and strength on each bottle. Stock keeps months in clean glass; weak working standards only a day or two.

The calibration curve

  1. Measure a blank (everything except the substance: the same water and reagents) and set the instrument to zero absorbance (or zero emission) with it.
  2. Measure 4–6 standards spread over the range you expect.
  3. Draw a graph on squared paper: concentration along the bottom, reading up the side. The points should lie on a straight line, or a smooth curve, through zero.
  4. Measure the unknown, find its reading on the side, go across to the line and down to read the concentration.
  5. Re-check one standard every 10 samples and at the end. If it has drifted more than 2–3 %, recalibrate and repeat the samples since the last good check.
00.10.20.30.40.50.6 012345 Concentration of standards (mg per litre) Absorbance reading of the unknown 2.4 mg/L blank standards do not read here: dilute and repeat
Fig. 1 A calibration curve. Five standards and a blank give points on a straight line. The unknown's reading is traced across to the line and down to its concentration. Only read concentrations within the range of the standards.

Never read off the end of the line. If the unknown is stronger than the strongest standard, dilute it and multiply the answer by the dilution.

Tip Measure everything twice, and include a "check sample" of known strength made up by someone else, without telling the person who measures. If it comes out wrong, so do the others. This is how a laboratory proves it can be trusted (The Scientific Method, Statistics and Probability).

The colorimeter

The simplest colorimeter is the eye, matching the sample against standards in matched tubes (The Medical Laboratory). The photoelectric colorimeter replaces the eye with a photocell, which is more exact and never tires.

Lamp 6 V, steady Lens parallel beam Filter colour absorbed Cuvette and lid 10 mm path Selenium cell current ∝ light Light-tight box matt black inside Battery Rheostat sets 100 with the blank Microammeter scale 0–100 % light µA
Fig. 2 The photoelectric colorimeter. A steady lamp shines through a lens, a coloured filter and the cuvette holding the sample onto a selenium photocell. The cell's current, read on a microammeter, falls as the sample absorbs more light.

What it is made of

Part What to use
Lamp 6 V, 3–5 W torch or car bulb (The Light Bulb), run from a lead-acid battery (The Rechargeable Battery) so it is steady
Lens Any small convex lens, 20–30 mm across, to make a parallel beam (Lenses)
Filters Coloured glass (Glass) or gelatin films dyed with clean dyes (Dyes and Mordants), one for each colour band
Cuvette A flat-sided glass cell with a 10 mm path, or matched test tubes
Detector A photovoltaic selenium cell 2–4 cm across (Photocells and Sensors)
Meter A microammeter or galvanometer reading 0–100 or 0–200 µA (Measuring Electricity)
Box Wood or sheet metal, blackened inside, with a lid over the cuvette

Building it

  1. Light path. Mount, in a row inside a blackened box: lamp; lens about one focal length from the lamp; a slot for the filter; the cuvette holder; then the photocell, close behind the cuvette. Total length 15–25 cm.
  2. Brightness control. A rheostat in the lamp circuit, or better an adjustable slit or iris in front of the cell, so you can set the reading to exactly 100 with the blank in place.
  3. Meter. Connect the selenium cell directly to the microammeter, with nothing else in the circuit. A low-resistance meter (under about 200 Ω) gives a current very nearly proportional to the light. Mark the dial 0–100 (transmittance).
  4. Cuvettes. Flat-sided cells are best: cement four strips of flat glass to a glass base with a waterproof cement, or blow and flatten them (Scientific Glassblowing and Glass-to-Metal Seals). Test tubes work if you choose well: fill twenty tubes with the same coloured solution, read each, and keep those that agree within 1 %. Scratch a mark on each so it always faces the same way.
  5. A light-tight lid over the cuvette is essential.

Better: the two-cell colorimeter. Lamp brightness wanders. Put a second cell where it sees the same lamp without the sample, and connect the two cells opposed through a galvanometer, with a slide-wire potentiometer to divide the reference cell's output. Balance the galvanometer to zero; read the answer off the slide-wire. Lamp changes then cancel out. This is the null-balance method of the potentiometer.

Using it: iron in drinking water

Iron above about 0.3 mg/L stains laundry and pipes and tastes metallic (Water Treatment for a Town).

  1. To 50 ml of sample add 1 ml of dilute Hydrochloric Acid (1 part strong acid to 1 part water) and 1 ml of 3 % Hydrogen Peroxide and Oxygen Bleach to turn all iron into its ferric form.
  2. Add 5 ml of 20 % potassium thiocyanate. It was made in chemical works by fusing yellow prussiate with sulfur and potash; that melt passes through free cyanide, so it is a works job, not a bench one (see the dangers in Prussian Blue and the Prussiates). A blood-red colour forms at once.
  3. Treat a blank (distilled water) and standards of 0.2, 0.5, 1, 2 and 3 mg/L iron exactly the same way.
  4. Read all within 15 minutes with a blue-green filter (about 480 nm). The colour fades after that.

The same instrument, with the right reagent and filter, measures copper (blue with ammonia, orange-red filter), manganese in steel (oxidised to purple permanganate, green filter, see Potassium Permanganate), phosphate in soil and water (blue with molybdate, see Molybdenum), haemoglobin and blood sugar (The Medical Laboratory) and chlorine left after disinfection.

The spectrophotometer

A filter passes a band 40–80 nm wide. A spectrophotometer uses a monochromator instead to pick a narrow band, 5–20 nm, of any colour. It is more exact, can tell two coloured substances apart in one solution, and draws a whole absorption spectrum, a fingerprint of the substance.

  1. Monochromator. The The Spectroscope with a second slit in place of the eyepiece. Light enters an entrance slit (0.2–1 mm), is spread by a prism or grating and focused onto an exit slit. Turning the prism with a screw and drum sweeps the colours across it; the colour that passes goes on through the cuvette to the detector.
  2. Wavelength scale. Calibrate the drum with known lines: sodium (589 nm), lithium (671 nm), hydrogen (656 and 486 nm) and mercury lamps (546 and 436 nm) (Arc Lamps, Neon and Fluorescent Lights).
  3. Detector. A narrow band carries little light. A selenium cell and meter may do for the middle of the spectrum with wide slits; for real work use a vacuum phototube and a valve amplifier (The Vacuum Tube, Amplifiers and Oscillators).
  4. Stray light, scattered inside the instrument, makes strong absorbances read too low. Blacken everything, add baffles, keep glass clean.
  5. Checking it. A piece of glass containing neodymium and praseodymium ("didymium" glass, see Rare Earths, Cerium and Phosphors) has sharp absorption bands that check the wavelength scale. A sealed solution of potassium dichromate (obtained by trade or from a chromate works, Chromium and Its Compounds) or copper sulfate, measured on the first day and kept dark, checks the absorbance scale.

Ultraviolet. Many colourless substances (proteins, medicines, the bases of DNA) absorb strongly at 200–350 nm. Ordinary glass blocks these wavelengths, so ultraviolet instruments need fused quartz prisms, lenses and cuvettes (Fused Quartz and Halogen Lamps) and a hydrogen discharge lamp. Build visible first.

The flame photometer

The flame photometer measures sodium, potassium, lithium and calcium, which have no good colour reactions, in blood, soil extracts, cement, glass and plant ash.

air Regulator and gauge, 0.7–1 bar Atomiser Sample Spray chamber fine mist goes up to the burner Drain and water trap big drops run away fuel gas Chimney with a window Flame steady, no draughts Lens Filter one element's colour Photocell Meter
Fig. 3 The flame photometer. Compressed air at a steady pressure draws the sample up a fine capillary and blows it into a mist. Large drops drain away; the fine mist feeds the burner. A filter passes only the colour of one element to the photocell.
  1. Air supply. A small compressor or a hand-pumped air tank (Bellows to Air Compressors), with a pressure regulator and gauge (Pressure and Vacuum Gauges) set to about 0.7–1 bar.
  2. Atomiser. Two glass tubes at right angles, as in a scent sprayer. Air blows from a jet about 0.5 mm across past the tip of a capillary about 0.3–0.5 mm across, which dips into the sample cup. The rushing air sucks the liquid up and tears it into mist (Scientific Glassblowing and Glass-to-Metal Seals).
  3. Spray chamber. A glass or plastic bulb about 200 ml. Big drops hit the wall and run out of a drain at the bottom through a U-trap of water (so gas cannot escape). Only fine mist, about 5 % of the sample, goes on.
  4. Burner. The mist mixes with fuel gas (coal gas, Natural Gas and Helium, or propane) and burns in a steady, non-luminous flame on a wide-mouthed burner with a chimney. The flame must not flicker: shelter it from draughts.
  5. Optics. A lens or concave mirror collects light from the flame and sends it through a filter to a selenium cell and microammeter, as in the colorimeter. Filters: yellow-orange for sodium, deep red for potassium (a dark red glass that passes only above about 700 nm), red for lithium. A monochromator in place of the filter separates them better.

Using it. Spray distilled water and set zero; spray the strongest standard and set full scale; then standards and samples, with water between. Working range is about 0–10 mg/L sodium or potassium.

  • Blood serum: dilute 1 part in 200 for sodium (normal 135–145 mmol/L becomes about 16 mg/L) and 1 in 20 for potassium.
  • Interference: a lot of sodium makes potassium read high, and calcium disturbs both. Make your standards resemble the samples: for serum potassium, add about the same sodium to the potassium standards.
  • Keep the air pressure, gas flow and sample viscosity the same for standards and samples. A sticky sample sprays slower and reads low.

The polarograph

The polarograph measures lead, zinc, cadmium, copper and nickel in water, foods and alloys down to about 1 mg/L, several at once.

How it works. A slowly rising voltage is applied between a tiny, constantly renewed mercury drop and a mercury pool, in the sample with plenty of potassium chloride. When the voltage is just enough to deposit one metal into the drop, the current rises in a step, a wave, and levels off. Where along the voltage the step comes tells you which metal; how high the step is tells you how much. In 0.1 mol/L potassium chloride, against a mercury pool, the steps come roughly at: copper near 0 V, lead about −0.4 V, cadmium −0.6 V, zinc −1.0 V. Calibrate your own with standards.

Building it

  1. Dropping mercury electrode. A thick-walled glass capillary with a bore of 0.05–0.08 mm, 6–10 cm long, joined by rubber or plastic tubing to a mercury reservoir 40–80 cm above it. Adjust the height until a drop falls every 3–5 seconds (Mercury).
  2. Cell. A small beaker or test tube with 10–20 ml of sample in 0.1 mol/L potassium chloride, and a pool of mercury at the bottom with a platinum wire sealed through the glass to connect it.
  3. Remove oxygen. Dissolved oxygen gives its own waves. Bubble pure nitrogen or hydrogen through the solution for 10 minutes (Liquid Air, Nitrogen and the Noble Gases, Hydrogen, Oxygen and Gases), then stop bubbling while measuring.
  4. Voltage. A 2 V cell across a slide-wire potentiometer turned slowly by hand or by a clockwork motor, sweeping from 0 to −2 V over 10–20 minutes.
  5. Current. A sensitive galvanometer reading 0.1–10 µA, with a shunt to change range and a large capacitor across it to damp the swings as each drop grows and falls (Measuring Electricity). Plot current against voltage by hand, or let a chart recorder draw it.

Liquid chromatography (HPLC, in outline)

Columns packed with finer particles separate far better than those of Chromatography, but need high pressure. High-performance liquid chromatography (HPLC) pumps solvent at 50–400 bar through a steel column about 4.6 mm across and 150–250 mm long, packed with silica particles of 3–5 µm, all the same size. A sample of about 20 µL is injected through a valve with a small loop. Each substance comes out at its own time, and a flow-through ultraviolet spectrophotometer cell, often at 254 nm, measures it as it passes. The height or area of each peak, compared with standards, gives the amount.

This is a late step: it needs uniform micro-particles of silica (Water Glass (Sodium Silicate)), piston pumps with ruby check valves (Synthetic Gems), stainless fittings (Alloy and Stainless Steels) and ultraviolet optics. Until then, a gravity-fed glass column with a flow-through colorimeter cell at its foot and a chart recorder gives real numbers.

The mass spectrometer

A mass spectrometer weighs atoms and molecules. It tells isotopes apart, finds leaks with helium (High-Vacuum Pumps and Gauges), and identifies gases and, behind a gas chromatograph, the substances in a mixture.

gas in Ion source hot filament; 70 V electrons Slits accelerate ions Magnetic field between pole pieces bends ions into half-circles: r = √(2mV/q) ÷ B light this mass heavy Exit slit one mass at a time Collector cup electrometer to pump 1–3 kV vacuum
Fig. 4 A 180° magnetic mass spectrometer. Electrons from a hot filament knock electrons off gas molecules, making ions. The ions are accelerated through slits by a high voltage, then bent into semicircles by a magnetic field: heavier ions follow wider circles. Only ions of one mass pass the exit slit into the collector.

How it works.

  1. Ionise. A trace of the gas leaks into a chamber under high vacuum. A hot tungsten filament (Tungsten and Filament Wire) sends electrons at about 70 V through it; they knock electrons out of the molecules, leaving them positively charged ions.
  2. Accelerate. A voltage V of 1–3 kV between slits speeds the ions into a narrow beam.
  3. Bend. In a magnetic field B they travel in circles of radius r = √(2 m V / q) ÷ B. Heavier ions bend less. For nitrogen (mass 28), 1,000 V and a field of 0.3 tesla, r is about 8 cm.
  4. Collect. An exit slit lets through only ions of one mass at a time into a small metal cup (a Faraday cup). The tiny current, 10⁻¹² to 10⁻⁹ amperes, is measured with an electrometer: a valve amplifier with a very high input resistance (The Vacuum Tube).
  5. Scan. Slowly change V (or B) and each mass in turn passes the slit. The current plotted against V is the mass spectrum: a row of peaks, each a mass, each height an amount.

What it takes. A vacuum of 10⁻⁵ millibar or better, from a diffusion pump (High-Vacuum Pumps and Gauges); an electromagnet with pole pieces giving an even field of 0.2–0.5 tesla across a gap of a few millimetres (The Electromagnet, Transformer Steel, Ferrite and Magnetic Shielding); a stable high-voltage supply (Power Supplies and Regulators); and an electrometer. A tube bent into a half-circle of 5–10 cm radius between the pole pieces, the 180° design of 1918, is the simplest. It is a hard but realistic project once high vacuum works, and its first job should be checking air: peaks at 28 (nitrogen), 32 (oxygen) in a ratio of about 4 to 1, and a small one at 40 (argon).

How to tell it worked

  • Standards fall on a straight line (or a smooth curve) through the blank.
  • Repeat readings of one sample agree within 1–2 % for a colorimeter, 2–3 % for a flame photometer.
  • Recovery test. Add a known amount of standard to a real sample. The measured increase should be 95–105 % of what you added. If it is not, something in the sample interferes.
  • A known check sample comes out right.

Note Troubleshooting. Readings drift → battery running down, photocell warming, or flame unsteady. Recheck blank and a standard often; use the two-cell circuit. Calibration curve bends over at the top → too concentrated, stray light, or too wide a band of colours. Dilute; use a narrower filter. Blank reads high → dirty cuvette or impure water or reagent. Flame photometer reads low and jumpy → capillary partly blocked. Clear it with a fine wire and spray clean water between samples. Polarograph waves have big spikes on top → oxygen left, or no damping. Bubble longer; a trace of gelatin in the solution suppresses the spikes.

Dangers

Danger Mercury vapour poisons slowly and invisibly. The polarograph uses a lot of it. Work over a tray that catches every spilled drop, keep mercury under water, never heat it, and recover all used mercury for cleaning (Mercury). Sweep up droplets with card and a brush, never a vacuum cleaner, and ventilate.

Danger Flame photometers burn gas next to bottles of sample and solvent. Never spray flammable solvents. Light the flame only after the air is flowing; turn the gas off first and the air last. Fit the drain trap, or gas will escape through it.

  • High voltage: a mass spectrometer's accelerating supply of 1–3 kV can kill. Interlock the covers (Wiring and Electrical Safety).
  • Reagents: never heat thiocyanate or prussiate reagents with strong acid; they can give off poisonous gas. Work by an open window or in a fume cupboard. Strong acids, see Sulfuric Acid and Nitric Acid.
  • Blood samples carry infection (The Medical Laboratory).

What this unlocks

Numbers that make Water Treatment for a Town, Chemical Fertilisers and steelmaking (Alloy and Stainless Steels) controllable, and medicine measurable (The Medical Laboratory, Intravenous Fluids). The mass spectrometer leads to helium leak detectors (Ion Pumps and Leak Detectors), isotope work for Nuclear Energy (Overview) and radiocarbon dating. Instruments that give a current also feed Instruments, Chart Recorders and Control Rooms, the control loops that hold a works steady (Process Controllers and Control Valves), and, later, The Computers that log thousands of readings.

How it was first done

Pierre Bouguer (1729) and Johann Lambert (1760) described how light fades through absorbing matter; August Beer linked it to concentration in 1852. Jules Duboscq's visual colorimeter of about 1854 was used for a century. Photoelectric colorimeters with selenium cells spread in the 1930s, and the Beckman DU spectrophotometer of 1941, with a quartz prism and a phototube, made ultraviolet measurement routine. Henrik Lundegårdh in Sweden developed practical flame photometry for plant analysis in the late 1920s; flame photometers for blood sodium and potassium were common by the 1950s. Jaroslav Heyrovský invented polarography in Prague in 1922, built an automatic recording polarograph with Masuzo Shikata in 1924–25, and won the Nobel Prize in 1959. J. J. Thomson separated neon isotopes with his positive-ray apparatus in 1912–13; Arthur Dempster built the 180° magnetic mass spectrometer in 1918 and Francis Aston the mass spectrograph in 1919. HPLC grew from the work of Csaba Horváth and others in the late 1960s.

Source: content/entries/chemistry/analytical-instruments.md