When someone cannot breathe enough, a machine can breathe for them until they recover. The hand-squeezed bag-valve-mask, the iron lung, the positive-pressure ventilator and CPAP built from simple parts, with the settings to use, how to watch the patient, and why too much pressure kills.

Difficulty
●●●●●
Time
A day to make a bag-valve-mask once you can mould rubber; a week for bubble CPAP; one to three months for a bellows ventilator or an iron lung, then long practice before using either on a patient
Steps from zero
61 prerequisites →

You need first

Medical OxygenAnaesthesia

This unlocks

Airway Equipment: Laryngoscopes and Tubes
The full path from nothing — 61 steps before this one
  1. Stone Tools
  2. Cordage and Rope
  3. Fire
  4. Finding and Preparing Clay
  5. Finding and Identifying Ores
  6. Sulfur
  7. How the Body Works
  8. Basic Woodworking
  9. Hunting and Trapping
  10. Charcoal
  11. Pottery
  12. The Kiln
  13. Examining a Patient
  14. Hides and Leather
  15. Salt
  16. Lime
  17. Bricks
  18. Mercury
  19. Fermentation
  20. Lye and Potash
  21. The Wheel and Axle
  22. Glass
  23. Firebrick and Refractories
  24. Bellows
  25. Distillation
  26. Saltpeter
  27. The Lathe
  28. Gears
  29. Iron (the Bloomery)
  30. Copper Smelting
  31. Laboratory Glassware
  32. Sulfuric Acid
  33. Pumps
  34. The Waterwheel
  35. Blacksmithing
  36. The Blast Furnace and Cast Iron
  37. Ether, Acetone and Solvents
  38. The Battery
  39. Magnets
  40. The Barometer
  41. Steel
  42. Sheet Metal and Rolling
  43. Anaesthesia
  44. The Steam Engine
  45. Bellows to Air Compressors
  46. Electrolysis
  47. Copper Wire
  48. Hydrogen, Oxygen and Gases
  49. Machine Tools
  50. Boilers and Pressure Vessels
  51. The Generator
  52. The Electric Arc Furnace
  53. Ammonia
  54. Ice and Refrigeration
  55. Calcium Carbide and Acetylene
  56. Welding
  57. Liquid Air, Nitrogen and the Noble Gases
  58. Metal Pipes and Tubes
  59. Pressure and Vacuum Gauges
  60. Pressure Vessels and Valves
  61. Medical Oxygen
  62. Breathing Machines and Ventilators

Do them roughly in this order. See it on the tech tree →

Why it matters

People stop breathing well enough for reasons that pass: deep Anaesthesia or an overdose, a newborn who does not start, a paralysing snakebite (Dangerous Plants and Animals), polio, tetanus, a crushed chest, severe pneumonia. If air is moved in and out of their lungs for hours, days or weeks, many will live. Mouth-to-mouth breathing (Emergencies) works for minutes; for longer you need a machine. From simplest to hardest:

  1. The bag-valve-mask, squeezed by hand: every clinic, every ambulance, every delivery room should have one.
  2. CPAP, which does not breathe for the patient but holds the lungs open: life-saving for babies and pneumonia, and cheap.
  3. The iron lung, which breathes for a paralysed patient by sucking on the outside of the chest.
  4. The positive-pressure ventilator, which blows air through a tube into the windpipe: the heart of intensive care.

Every one of them can kill by over-inflating the lungs.

How it works

You breathe in by pulling the chest wall out and the diaphragm down, lowering the pressure in the lungs slightly; air flows in. Breathing out is passive: the springy lungs recoil. A machine can copy this two ways:

  • Negative pressure: lower the pressure around the body (the iron lung). The chest expands just as in natural breathing.
  • Positive pressure: raise the pressure at the mouth or windpipe and blow air in (the bag, the ventilator). Breathing out is still passive.

Breathing pressures are tiny, so they are measured in centimetres of water (cmH₂O): the pressure that holds up a water column that many centimetres tall in a U-tube. 1 cmH₂O is about 0.74 mmHg. A normal breath needs only 5–10 cmH₂O. Stiff, sick lungs may need 20–30.

The size of one breath is the tidal volume: about 6–8 ml per kg of body weight, roughly 450–550 ml for an adult, 25 ml for a 3 kg baby. Breaths a minute times tidal volume is the minute volume, about 6 litres a minute for a resting adult.

PEEP (positive end-expiratory pressure) means keeping a small pressure, usually 5 cmH₂O, in the lungs even at the end of breathing out. Sick lungs have millions of tiny air sacs that collapse when emptied and are hard to reopen; PEEP keeps them open. CPAP is the same steady pressure for a patient who still breathes for themselves.

What you need

Item For See
Natural rubber, latex, moulds Bags, masks, valves, tubes Rubber
Leather or rubberised cloth Bellows Hides and Leather, Bellows
Oxygen with flowmeter Most patients Medical Oxygen
Glass or metal tubes, T-pieces Circuits Laboratory Glassware, Metal Pipes and Tubes
Tall jars, water PEEP and safety valves, manometer Pressure and Vacuum Gauges
Electric motor with gearbox, or a hand crank Ventilator, iron lung The Electric Motor, Gears
Plywood and battens, or sheet steel Iron lung Plywood, Veneer and Boards, Sheet Metal and Rolling
Pulse oximeter Watching every patient Pulse Oximeters and Patient Monitors
Suction (a bulb or foot pump) Clearing the airway Airway Equipment: Laryngoscopes and Tubes

The bag-valve-mask

Oxygen reservoir 1.5–2.5 litres oxygen in, 10–15 L/min Inlet valve lets air in only squeeze Self-inflating bag rubber 3–4 mm thick Pop-off valve opens at 40 cmH₂O to patient Patient valve one-way duckbill breath out Mask air-cushion rim fittings: 22 mm mask, 15 mm tube
Fig. 1 The bag-valve-mask taken apart. Squeezing the self-inflating bag pushes air through the patient valve into the mask; the patient breathes out through the side port. The bag springs back and refills through the inlet valve, from the oxygen reservoir if one is fitted. A pop-off valve on small bags limits the pressure.

How it is built

A bag-valve-mask (BVM, also called a self-inflating bag or "Ambu bag") has five parts:

  1. The bag, an oval of thick, springy rubber that pops back into shape when released, refilling itself with air.
  2. The inlet valve at the back: a thin rubber flap over holes, which lets air into the bag but not out.
  3. The patient valve at the front (a non-rebreathing valve): when you squeeze, a soft "duckbill" or flap opens toward the patient and at the same time closes the side port; when you let go, the patient's breath pushes out through the side port and never goes back into the bag.
  4. The mask, a soft rubber cup with an air-filled rim that seals over nose and mouth.
  5. The oxygen reservoir: a thin bag of 1.5–2.5 litres on the inlet, filled by oxygen at 10–15 litres a minute, so each refill is nearly pure oxygen. Without it, oxygen piped into the bag gives only about 40–50 %.

Small bags for children and babies carry a pop-off valve: a spring-loaded or weighted disc that lets air escape above about 40 cmH₂O, protecting small lungs.

Making one

  • Bag: dip a smooth oval former of fired clay or wood into compounded latex again and again (Rubber) until the wall is 3–4 mm thick, vulcanise it, and peel it off through the neck. It must spring back fully within a second of being squashed flat. Old bags were moulded from sponge rubber with a thin outer skin; that works too.
  • Valves: cut flaps from thin rubber sheet (0.3–0.5 mm), cemented at one edge over a disc drilled with five or six 4 mm holes. A duckbill is a rubber cone dipped on a pointed former and slit across the tip. Each valve must open with the faintest puff and close tight against the reverse.
  • Mask: dip a rubber cup on a former shaped like the lower face, then glue a hollow rubber ring round the rim and inflate it slightly.
  • Joints: standard sizes save lives, because every mask must fit every bag and later every tube. Use a 22 mm outside diameter for mask fittings and 15 mm for tube fittings, as the world has done since the 1960s.
Size Bag volume Breath to give Rate
Adult 1.5–1.6 litres squeeze about one-third (500 ml) 1 every 6 seconds
Child 0.5 litres until the chest just rises 1 every 2–3 seconds
Newborn 0.25 litres until the chest just rises 40–60 a minute

Using it

  1. Tilt the head back and lift the chin (not in a suspected neck injury: lift the jaw only). Clear the mouth.
  2. The E–C grip: your thumb and forefinger make a "C" pressing the mask down; your other three fingers make an "E" under the bony jaw, lifting it into the mask. Seal is everything. With two rescuers, one holds the mask with both hands and the other squeezes.
  3. Squeeze over 1 second, just until the chest rises. Let go fully and let the chest fall.
  4. Add oxygen at 10–15 litres a minute with the reservoir. For a newborn born at full term, start with plain air: pure oxygen harms them.

Good or bad: a good bag refills in under a second, every valve clicks cleanly, and when you block the patient end and squeeze, air escapes only through the pop-off (on small bags) and nowhere else. A bag that stays dented, valves that stick, or a mask that leaks round the nose is useless in an emergency. Check every bag every week.

CPAP from simple parts

Two flowmeters air O₂ mixed, 5–8 L/min Humidifier warm water, 37 °C Soft nasal prongs snug, skin not white Breathed-out gas dips into water 5 cm = 5 cmH₂O steady bubbling = circuit sealed
Fig. 2 Bubble CPAP for a baby. Blended oxygen and air flow through a humidifier to soft nasal prongs. The breathed-out gas leaves through a tube dipped 5 cm into water: the depth sets the pressure in the baby's lungs, and steady bubbling shows the circuit is sealed.

Bubble CPAP is the most useful breathing machine in this chapter: it costs little, needs no electricity beyond the oxygen supply, and saves babies with stiff, sticky lungs (common in those born early), and children with severe pneumonia or bronchiolitis. A large trial in Bangladesh found it saved more children with severe pneumonia than oxygen alone.

  1. Gas: air and oxygen from two flowmeters, mixed, at 5–8 litres a minute for a baby. For babies born early aim for an oximeter reading of 90–95 %; too much oxygen blinds them (Medical Oxygen).
  2. Humidify and warm it: bubble it through warm water at about 37 °C.
  3. Nasal prongs: two short soft rubber tubes that fill the nostrils snugly without whitening the skin, on a small manifold, held by a cap or ties.
  4. The pressure: run the breathed-out limb (a tube about 1 cm bore) into a bottle of clean water, the tube's end 5 cm under the surface. The pressure in the baby's lungs is 5 cmH₂O, set purely by the depth. Mark the tube in centimetres. Start at 5, go up to 7–8 if the baby still struggles.
  5. It is working when the water bubbles steadily. No bubbles means a leak, usually the mouth open: close it with a soft chin strap or a dummy.
  6. Vent the stomach: CPAP pushes air into the stomach; pass a soft feeding tube (Intravenous Fluids) and leave its end open.

Adult CPAP needs far more flow, at least 60 litres a minute, because adults draw air in fast. Use a blower: a fan driven by a motor, like a vacuum cleaner running slowly, feeding a tight mask with a few small vent holes to flush out breathed air. Set the pressure (5–10 cmH₂O for pneumonia, 4–15 for people who stop breathing in their sleep) by the motor's speed, and check it with a water manometer.

The iron lung

air in Head outside breathes normally Soft rubber collar seals round the neck Portholes for nursing hands 12–15 cm, rubber sleeves Water gauge Sliding bed Tank, sealed about 2.1 m × 65 cm Bellows 15–20 litre stroke Motor and crank, 14–20/min hand lever for power cuts
Fig. 3 The iron lung. The patient lies on a sliding bed inside a sealed tank, head outside through a soft rubber collar. A motor-driven bellows under the tank lowers and raises the pressure inside; as it falls, the chest expands and air flows in through the nose and mouth. Portholes let nurses reach in.

The iron lung needs no tube in the windpipe. It suits patients who are awake and can swallow, but whose breathing muscles are paralysed by polio, snakebite or nerve disease, for weeks or for life.

Building one:

  1. The tank: a box or cylinder big enough for an adult from neck to feet, about 2.1 m long and 60–70 cm across. Steel (as most were) or plywood 18–20 mm thick with battens every 30 cm: at 20 cmH₂O below outside pressure, the air presses about 200 kg on every square metre of wall, and a weak box bulges inward and leaks.
  2. The head end is a removable plate with a hole for the neck. The collar is a sheet of soft sponge rubber with a hole about two-thirds of the neck's size, stretched over the neck to seal without choking.
  3. The bed: a padded tray that slides out on rails, so the patient can be loaded and taken out quickly.
  4. Portholes: four to six round holes, 12–15 cm across, along the sides, each closed by a rubber sleeve or a hinged glass lid, for nursing hands. A window to see the patient, and a mirror above the head so the patient can see the room.
  5. The bellows: a leather or rubberised cloth bellows under the tank, driven through a crank by a geared motor. Its stroke must be 15–20 litres: much of it only stretches the air in the tank itself. Make the stroke adjustable by moving the crank pin, and the rate by changing pulleys.
  6. A pressure gauge (a water U-tube) on the tank, and a hand lever on the bellows for power cuts. When the power fails, people take turns pumping, all night if needed.

Settings: for an adult, 14–20 breaths a minute with the tank 15–25 cmH₂O below outside pressure at the deepest point, adjusted until the chest moves comfortably, the patient can talk in short sentences on each breath, and the oximeter reads above 94 %. Children need less pressure and more breaths.

A cuirass is a lighter cousin: a rigid shell over the chest and belly only, sealed at the edges, connected to the same bellows. It is enough for patients who can breathe partly by themselves.

The mechanical ventilator

A positive-pressure ventilator blows air into the lungs through a tube in the windpipe. It is far more powerful than an iron lung: it can inflate stiff, pneumonia-filled lungs, it lets nurses turn and clean the patient, and the cuff on the tube keeps vomit and saliva out of the lungs. It also needs an anaesthetist's skills, drugs to keep the patient calm, constant nursing, and a reliable oxygen supply. It is a tool of a real hospital (Running a Hospital).

The endotracheal tube

The endotracheal tube is a curved tube of smooth rubber or plastic passed through the mouth, between the vocal cords, into the windpipe. Near its tip a thin cuff, a balloon blown up through a fine side tube, seals against the windpipe wall.

  • Sizes: adult women 7.0–7.5 mm inside bore, men 7.5–8.0 mm, cut so that about 21–23 cm sits at the front teeth. For a child over one year, bore in mm ≈ age ÷ 4 + 4 (without a cuff).
  • Cuff pressure: 20–30 cmH₂O, checked with the manometer. Higher pressure cuts off the blood supply of the windpipe lining.
  • Making it: dip latex on a curved polished rod, with a fine wire laid along it before the last coats; pull the wire out to leave the cuff's inflation channel. Add the cuff as a thin band dipped over a strip dusted with talc, so it does not stick and can balloon.

Putting the tube in (intubation), using a lighted blade called a laryngoscope to see the vocal cords (Airway Equipment: Laryngoscopes and Tubes), is learned only by supervised practice. A tube in the gullet instead of the windpipe kills within minutes: check that both sides of the chest rise and that breath sounds are heard over both lungs, not gurgling over the stomach (Examining a Patient). For more than about a week, a surgeon places a tracheostomy tube through the front of the neck.

A bellows ventilator you can build

Reservoir bag oxygen in inlet valve Bellows up to 1 litre Motor Crank pin position sets volume, speed sets breaths a minute 40 cm Safety valve pressure can never pass 40 cmH₂O Manometer 0–60 cmH₂O Patient valve one-way to tube Swedish nose keeps the airway moist PEEP valve tube 5 cm deep = 5 cmH₂O kept in lungs breath out
Fig. 4 A bellows ventilator. A crank drives a bellows that fills from an oxygen-enriched reservoir and pushes each breath through a one-way patient valve to the tube. Breathed-out gas leaves through a tube dipped 5 cm into water, the PEEP valve. A second tube dipped 40 cm deep is the safety valve: if pressure rises above 40 cmH₂O, gas bubbles out there instead of bursting the lungs.
  1. Bellows: a concertina of rubberised cloth between two boards, up to 1 litre per stroke for adults, about 200 ml for children; a separate small bellows for babies.
  2. Drive: an electric motor through a worm gearbox turning a crank, 10–30 turns a minute. The crank pin slides in a slot, so its distance from the centre sets the stroke: the tidal volume. Mark the slot in millilitres after measuring (below).
  3. Gas in: the bellows fills through a one-way valve from an open-ended reservoir bag fed with oxygen. More oxygen flow, more oxygen in each breath.
  4. Patient valve: the same non-rebreathing valve as the bag-valve-mask. Breath goes in; breathed-out gas leaves by the side port.
  5. PEEP valve: the side port's tube dips into a jar of water, its end 5 cm below the surface for 5 cmH₂O of PEEP.
  6. Safety valve: a T-piece on the line to the patient, its tube dipped 40 cm into a tall jar of water. Nothing the crank does can push the lungs above 40 cmH₂O; the excess bubbles out. This is the most important part of the machine.
  7. Manometer: a U-tube of water, 0–60 cmH₂O, on the patient line, where the nurse can see it.
  8. Humidity: dry gas through a tube dries the airway and plugs it with thick mucus. Put a "Swedish nose" (a short canister of rolled paper or wool that holds the warmth and moisture of each breath out and gives it back) between the tube and the valve, or bubble the gas through warm water.
  9. Disconnection alarm: a float in the manometer touches a contact each breath; if no breath reaches 8 cmH₂O within 15 seconds, a battery bell rings (Electric Bells, Buzzers and Alarms). Or use a pressure sensor and Microcontrollers.

Calibrating the volume: connect the machine to a thin bag or an inverted, water-sealed bell jar marked in 100 ml steps (a spirometer) and mark the stroke for each 100 ml.

Settings

Setting Adult start Notes
Tidal volume 6–8 ml per kg of ideal weight (about 450 ml) Ideal weight, not actual: the lungs do not grow with fat
Rate 14–20 a minute (children 20–30, babies 30–40) Adjust to keep the patient calm and pink
Time in : time out 1 : 2 Breathing out must finish before the next breath
PEEP 5 cmH₂O Up to 10–12 in severe pneumonia with low oxygen
Oxygen Start high, then lower Keep the oximeter at 92–96 %
Peak pressure Keep below 35 cmH₂O Safety valve at 40

Watching the patient

Write down every hour: peak pressure, rate, volume, oxygen flow, oximeter, pulse and blood pressure (Pulse Oximeters and Patient Monitors), chest movement and colour. A sudden rise in peak pressure means a blocked tube (suck it out), a kinked tube, the patient fighting the machine, or a burst lung (below). A sudden fall means a leak or disconnection.

Every day, ask whether the cause is improving. When the patient is awake, coughing, and needs little oxygen and PEEP of 5 or less, try 30–120 minutes of breathing through the tube with only humidified oxygen; if they stay comfortable, take the tube out.

How to tell it worked

  • The chest rises gently and equally on both sides with each breath, and falls fully.
  • Colour turns pink and the oximeter rises above 92–94 %.
  • The pulse slows toward normal as oxygen improves.
  • The machine: blocked at the patient end, the manometer climbs to 40 and the safety jar bubbles; disconnected, the alarm rings within 15 seconds. Test both before connecting any patient.
  • Volume: the spirometer shows the set volume within about 10 %.

Note Troubleshooting. Chest does not rise with the bag → poor seal (two-person grip), head not tilted, or something in the throat. Belly swells → air is going into the stomach: breathe more gently, slower, and pass a stomach tube. Bubble CPAP not bubbling → mouth open or prongs too small. Iron lung patient pink but tired, pressure won't fall → leak at the collar or portholes. Ventilator pressure creeping up over hours → mucus in the tube: humidify better, suck it out.

Dangers

Danger Over-inflation bursts lungs. Too much pressure or volume tears the air sacs. Air leaks around the lung (pneumothorax) and, under positive pressure, builds up until it squashes the lung and the heart (tension pneumothorax). Signs: sudden rise in airway pressure, falling oximeter and blood pressure, one side of the chest not moving and sounding hollow when tapped, windpipe pushed to one side. It kills within minutes. A trained person lets the air out with a wide needle through the chest wall on the affected side, in the second space between the ribs, in line with the middle of the collarbone (in adults many now prefer the fourth or fifth space, just in front of the middle of the armpit, where the chest wall is thinner), then places a chest drain (Basic Surgery). Prevent it: small breaths, safety valves, the pop-off, and never squeezing a bag "as hard as you can".

Danger Disconnection or power failure kills in minutes. Keep a bag-valve-mask at every ventilated bedside, a hand lever on every iron lung, and someone awake beside every patient on a machine, always.

  • Breathing too fast traps air: the chest cannot empty, pressure builds and the blood pressure falls. Let each breath out finish.
  • Positive pressure lowers blood pressure by squeezing blood out of the chest veins. Give fluids (Intravenous Fluids) and use the smallest pressure that works.
  • Stomach inflation above about 20 cmH₂O at the mouth causes vomiting, and vomit breathed into the lungs.
  • Infection: tubes carry germs into the lungs. Sterilise all tubing between patients (Sterilising Instruments and Dressings), keep the head of the bed raised 30°, clean the mouth every few hours.
  • Oxygen fire: no flames or sparks near oxygen (Medical Oxygen).
  • The patient is a person. Paralysed patients in iron lungs are awake and frightened. Talk to them, explain everything, and nurse their skin, eyes and bowels (Nursing the Sick).

What this unlocks

Ventilators make long operations with muscle-relaxing drugs safe, and with Pulse Oximeters and Patient Monitors, Medical Oxygen and Kidney Dialysis they form the intensive care unit of a hospital. The same valves and pressures serve Diving and Underwater Work gear and oxygen for high-flying Gliders and Aeroplanes.

How it was first done

In 1543 Andreas Vesalius described keeping an animal alive by blowing into its windpipe through a reed. London's Royal Humane Society promoted bellows to revive the drowned from the 1770s, until in 1827 Jean-Jacques Leroy d'Étiolles showed that bellows could burst the lungs.

Philip Drinker and Louis Agassiz Shaw at Harvard built the first practical electric iron lung in 1928. John Emerson made a cheaper, quieter one in 1931, and in the late 1930s Edward Both in Australia designed a plywood version, of which Lord Nuffield gave many hundreds to hospitals across the British Empire.

In the 1952 polio epidemic in Copenhagen, with far too few iron lungs, the anaesthetist Bjørn Ibsen proposed tracheostomy and hand-squeezed positive-pressure breathing; about 1,500 medical and dental students squeezed bags in shifts for weeks. Deaths among paralysed patients fell from nearly 90 % to about 40 %, and the modern intensive care unit grew from that effort. In 1953–56 Henning Ruben and Holger Hesse in Denmark made the self-inflating Ambu bag with its non-rebreathing valve. George Gregory in San Francisco used CPAP for premature babies in 1971, and Colin Sullivan in Sydney treated sleep apnoea with a CPAP blower built from a vacuum-cleaner motor in 1981.