Radiotherapy (Overview)
Radiation kills cancer cells a little more readily than it kills healthy ones, and that small difference, spread over daily doses for weeks, cures real cancers. How it works, the X-ray therapy machine and the treatment couch, how a dose is planned, measured and shielded, and why an abandoned radioactive source is one of the deadliest objects a scavenger can find. Principles and safety only.
- Difficulty
- ●●●●●
- Time
- Years. A society needs diagnosis, pathology, dosimetry and a trained team before a single patient can be treated safely
- Steps from zero
- 46 prerequisites →
This unlocks
—The full path from nothing — 46 steps before this one
- Fire
- Cordage and Rope
- Stone Tools
- Finding and Preparing Clay
- Finding and Identifying Ores
- Sulfur
- Hunting and Trapping
- Basic Woodworking
- The Kiln
- Pottery
- Charcoal
- Hides and Leather
- Lime
- Mercury
- Lye and Potash
- Fermentation
- The Wheel and Axle
- Glass
- Bellows
- Saltpeter
- The Lathe
- Copper Smelting
- Iron (the Bloomery)
- Lead
- Laboratory Glassware
- Sulfuric Acid
- Pumps
- Static Electricity and the Leyden Jar
- Lenses
- Blacksmithing
- Gold and Silver
- Nitric Acid
- The Battery
- The Vacuum Pump
- Silver Nitrate and Silver Salts
- The Light Bulb
- Copper Wire
- The Electromagnet
- The Vacuum Tube
- Mirrors
- The Induction Coil
- Detecting Radiation
- Vacuum Coating and Thin Films
- X-Rays
- High-Vacuum Pumps and Gauges
- Particle Accelerators (Overview)
- Radiotherapy (Overview)
Do them roughly in this order. See it on the tech tree →
Why it matters
About one person in three develops a cancer at some point in life, and roughly half of all cancer patients benefit from radiation treatment at some stage. For some cancers radiation alone cures: early cancer of the skin, the larynx, the cervix, the lip, some lymphomas. For many others it is what makes surgery work — shrinking a tumour before the knife, or killing the cells left behind. And for people past cure it is the most effective relief there is: one treatment can take away the pain of cancer in a bone within days, or free a spinal cord being crushed by a tumour before paralysis becomes permanent.
It is also the most dangerous machine in a hospital. Radiotherapy delivers, on purpose, doses ten to twenty times the whole-body dose that would kill the patient outright, aimed at a small volume. Get the aim, the dose or the arithmetic wrong and you injure or kill the person you are treating, and the injury appears weeks or months later when nothing can be undone.
Warning This chapter is principles and safety only. It explains how radiation treats cancer, what the machines and their furniture are, and how dose, planning and shielding work, so that a rebuilt society can judge what it would take and can recognise the hazards left behind by the old one. It deliberately gives no method for preparing, extracting, concentrating, loading or handling a radioactive source. That work killed many of the people who pioneered it, and no description in a book can make it safe. The achievable radiotherapy machine for a rebuilding society is an X-ray machine — a tube you switch off when you leave the room.
How radiation treats cancer
Radiation does its damage by ionisation: an X-ray or gamma ray tears electrons loose, those electrons break chemical bonds, and the bonds that matter are in DNA (Heredity and Genetics). Most of the damage is indirect — the radiation splits water molecules into violently reactive fragments, and those attack the DNA nearby. A double-strand break that the cell cannot repair kills it, usually at its next attempt to divide.
Why does this help, when a tumour is made of the patient's own cells?
- Dividing cells die more easily. Cancer cells divide constantly; most healthy adult tissue divides slowly or not at all.
- Healthy tissue repairs better. Give the dose in small daily portions and normal cells mend most of their damage overnight, while tumour cells, whose repair machinery is often faulty, do not. This is fractionation, the single most important idea in the subject.
- The tumour's survivors become more vulnerable as it shrinks and oxygen reaches cells that had none. Well-oxygenated cells are two to three times easier to kill than starved ones.
Against this: the curing dose must be given through healthy tissue, every organ has a limit, and a few cancers are simply radiation-resistant. The art is finding a path through the body that delivers a killing dose to the tumour while every normal structure on the way stays under its own limit.
Dose, and the numbers that matter
Dose is measured in grays (Gy): 1 gray is 1 joule of energy absorbed per kilogram. It is a tiny amount of energy — 2 Gy would warm tissue by about half a thousandth of a degree — and it is lethal to cells because of where the energy goes.
| Dose | |
|---|---|
| Natural background, a year | 0.002–0.003 Gy |
| Whole body, kills about half of those exposed | 4–5 Gy |
| One daily treatment fraction | 1.8–2 Gy |
| Curative course, most solid tumours | 60–70 Gy in 30–35 daily fractions over 6–7 weeks |
| Curative course, skin cancer (small field) | 45–55 Gy in 20 fractions, or 35 Gy in 5 |
| Pain from cancer in bone | 8 Gy in one fraction, or 20 Gy in 5 |
| Half the dose is given, half the cure | There is no half-cure: an interrupted course often achieves nothing |
Limits of normal tissue (total dose, in 2 Gy fractions, above which serious permanent harm becomes likely):
| Organ | Limit |
|---|---|
| Lens of the eye | 5–10 Gy (cataract) |
| Testis or ovary | 2–3 Gy (permanent sterility) |
| Whole kidney | 23 Gy |
| Lung, whole | 18–20 Gy |
| Spinal cord | 45–50 Gy — exceed it and the patient is paralysed, months later |
| Skin | 30 Gy before breakdown becomes likely |
Treatment is five days a week, not seven: the weekend break is part of the design. Missing further days lets the tumour regrow, and a course stretched out by interruptions loses much of its effect.
The radiotherapy machine
Four kinds of machine have been used. Only the first two are realistic for a rebuilding society, and they are the same machine at two voltages.
Superficial and orthovoltage X-ray therapy: the achievable machine
A radiotherapy machine of this type is an X-Rays machine built for continuous output rather than for an instant snapshot:
- A hot-cathode X-ray tube with a thick water- or oil-cooled tungsten target (Tungsten and Filament Wire), rated to run for minutes at a time at 10–20 mA — the hard part, since a diagnostic tube is rated for fractions of a second. Oil immersion in an earthed metal tank with a thin window.
- A high-voltage supply at 50–150 kV (superficial therapy, for skin) or 200–300 kV (orthovoltage or "deep" therapy), smoothed and stable: the dose rate changes with the voltage, so a wandering supply means an unknown dose (Power Supplies and Regulators, The Transformer).
- Filtration. A raw X-ray beam is a mixture of energies, and the softest rays never get past the skin — they only burn it. A metal filter at the tube's mouth absorbs them: 0.5–3 mm of aluminium for superficial work, 1–4 mm of copper backed by 1 mm of aluminium for orthovoltage. The filter is part of the prescription. Measure the beam's penetration as its half-value layer — the thickness of copper or aluminium that halves its intensity — and write it on the machine.
- An applicator: a lead-lined metal cone or box, which fixes the focus-to-skin distance (15–30 cm for superficial, 30–50 cm for orthovoltage) and defines the treated field exactly, its open end resting on or just off the skin. A set from 2 × 2 cm up to 15 × 15 cm covers most work. Light shining down it from a lamp and mirror shows the field on the skin.
- Two independent timers, each able to shut the beam off alone, plus a milliammeter and a voltmeter recorded for every treatment.
- Interlocks: door switch, warning lamp inside and out, a beam-on lamp visible from the couch, and emergency-off buttons at the control desk and inside the room. Particle Accelerators (Overview) describes this discipline in full, and it applies at 250 kV just as at 6 MV.
Orthovoltage is genuinely useful: it cures skin cancers, treats superficial lymph nodes and relieves bone pain. Its limits are brutal — the dose at the skin is always higher than at the tumour, and bone absorbs two to three times more than soft tissue at these energies, so ribs and jaws are injured and tumours behind bone cannot be reached.
The megavoltage machines
A linear accelerator making 4–20 MV X-rays is the modern machine, and Particle Accelerators (Overview) covers its construction, vault and interlocks. It needs a megawatt microwave source, copper machined to thousandths of a millimetre, a high vacuum and a trained physics team — years of industry away, but the right target to aim at, because megavoltage beams spare the skin, pass bone evenly and reach deep tumours.
Cobalt-60 and radium machines belong to history and to the hazard section below. A cobalt teletherapy unit is mechanically simple — a shielded head holding an intensely radioactive source, with a shutter — and that simplicity is a trap: its source cannot be switched off, decays whether used or not (losing half its strength in 5.3 years), and stays lethal for decades after the machine is scrapped. Radium was used in such machines and as sealed tubes and needles placed into or beside a tumour (brachytherapy), which cured cancer of the cervix when nothing else could, and which irradiated the hands that handled the sources.
The treatment couch
The treatment couch is not furniture; it is a measuring instrument. If the patient's tumour is not in the same place relative to the beam today as it was yesterday, the week's arithmetic is worthless.
- Top: flat, 190–200 cm long and 50 cm wide, rigid enough to sag no more than 2 mm at the middle under a 120 kg patient. Make it as a torsion box of thin plywood over a frame (Plywood, Veneer and Boards, Doors, Windows and Joinery) or as a welded steel frame with a thin deck. Thin and low-absorbing is the point — a beam from below must pass through the couch. A removable panel over a mesh of stretched nylon or linen cords, like a tennis racket, lets a beam through from underneath almost unattenuated.
- Movements: height 70–120 cm, travel along and across of at least 20 cm each, each with a scale and a positive lock. Set and read positions to 1 mm. A screw jack (Screws, Nuts and Bolts, Simple Machines) under a parallel linkage gives a steady height drive; two dovetail slides give the horizontal movements.
- Alignment marks. Fix two thin light beams, one vertical and one horizontal, crossing where the beam axis meets the couch's reference height. Before lasers this was done with a plumb bob hung from the tube housing and a front pointer — a graduated rod that slides in the applicator and touches the skin at a known distance. Both still work and are easy to make.
- Patient marks. Mark the field edges and centre on the skin with gentian violet or indian ink; for a long course, three tiny permanent tattoo dots pricked in with a needle are the reliable way (Ink).
The immobilisation shell
An immobilisation shell holds a patient — above all one treated about the head and neck, where a 5 mm error can mean a dose to the spinal cord — in the same position every day. Make it from plaster of Paris bandage laid over the face and shoulders with breathing and eye holes cut, built up to 4–5 mm thick, reinforced with a second layer and mounted on a board that indexes into the couch (Gypsum and Plaster of Paris). It reproduces the position to about 3–5 mm, against 10–15 mm for a patient simply lying still. Line it with soft felt, and remake it when the patient loses weight — which they will.
Shielding blocks
A shielding block is a slab of lead, shaped to the outline of what must be protected, mounted on a thin tray at the mouth of the applicator. At 250 kV, 3 mm of lead cuts the beam to a tenth and 10 mm to about a thousandth, so blocks are light and easily sawn. Cast them in plaster or sand moulds from scrap lead (Lead, galena is the ore), file the edges, and support them on 2 mm aluminium or thin plywood that clips to the applicator, located so a block cannot be fitted the wrong way round. Common uses: an eye shield, a block over the testes, a cut-out that turns a square field into the shape of the lesion.
Measuring the dose
This is the step that separates radiotherapy from assault. The machine's milliamps and minutes tell you nothing about grays until somebody has measured the relation.
The thimble ionisation chamber
A thimble ionisation chamber is a small air-filled cavity — typically 0.3 to 0.6 cm³, with a wall of graphite-coated plastic or thin carbon about 0.5 mm thick and a fine central electrode of aluminium or graphite. Radiation ionises the air inside; a polarising voltage of 150–300 V sweeps the ions to the electrodes and the tiny current — a few picoamps — is measured by an electrometer, in essence the vibrating-reed or valve amplifier of Electronic Test Equipment and Detecting Radiation built for extremely small currents. The charge collected per gray is fixed by the chamber's volume and by air's known energy per ion pair (about 34 joules per coulomb of charge liberated in air), which is what makes absolute measurement possible at all.
The practical procedure:
- Calibrate the chamber against a standard. A standards laboratory does this with a large free-air chamber in which the whole electron track is contained and every quantity is measured directly. Without one, you can build a free-air chamber — big, fussy, and only workable below about 300 kV — or you can do without absolute dosimetry and not treat patients. There is no honest third option.
- Measure the machine's output in grays per minute, at a stated depth in a water phantom (a tank of water, which absorbs radiation very nearly as tissue does; see Ultrasound, CT and MRI) for each applicator and filter combination, and tabulate it.
- Measure the depth dose: move the chamber down through the water in 1 cm steps and record the dose as a percentage of its maximum. These percentage depth dose tables, plus the field size and distance, are what every treatment calculation uses.
- Check the output every treatment day with the same chamber in the same jig, and plot it. A 5% drift means stop and find out why; a 20% error means somebody is harmed.
- Cross-check with a second method — film darkening against a step wedge (Making Photographic Papers and Plates, X-Rays), or better a chemical dosimeter in which radiation converts ferrous iron to ferric in acid solution, measured by colour (The Spectroscope). Chemical dosimetry needs 10 Gy or more to read, which suits therapy exactly.
Planning a treatment
- Know what you are treating. Radiotherapy without a tissue diagnosis under a microscope (Tissue Slides and the Microtome) and without imaging to find the tumour's extent (X-Rays, Ultrasound, CT and MRI) is guesswork that irradiates healthy people. Make these first.
- Draw the patient. An outline of the body at the level of the tumour, traced with a flexible lead wire, with the tumour and the organs at risk drawn in.
- Choose the beams. A single field will do for a skin lesion. For anything deeper, use a parallel opposed pair — identical fields from front and back — which makes the dose through the middle of the patient far more even than one field can. Three or four fields from different angles concentrate it further and spare the skin more.
- Add the fields up by hand at a grid of points through the patient, using the depth-dose tables. The results were drawn as isodose curves — contours of equal dose — on the body outline.
- Have a second person repeat every calculation independently before the first treatment. This rule is written in the deaths of patients overdosed by arithmetic errors and, later, by faulty machine software.
- Simulate. Set the patient up exactly as for treatment, under a diagnostic X-ray tube at the same geometry, and take a film to prove the field covers the tumour and misses the cord. Mark the skin from that session.
How to tell it worked
Daily, on the machine: the output check against its plotted history; the timer's accuracy (compare timed runs of 1 and 2 minutes — the dose should be exactly double, and if it is not, the shutter has a lag to allow for); the coincidence of the light field with the radiation field, checked by exposing a film at the applicator's mouth with pins pushed through at the light field's corners (they must land on the edge of the darkening within 2 mm); interlocks tested by opening the door mid-beam.
In the patient, over weeks: the tumour shrinks — a skin lesion crusts and flattens, a lump becomes smaller and harder, pain in a bone eases within 1–3 weeks. Expected reactions: the skin in the field reddens at 2–3 weeks like sunburn, hair in the field falls out, a treated mouth or throat becomes sore and dry, a treated belly brings nausea and loose stools. Not acceptable: a field that breaks down into a wet ulcer before the course is finished (dose or filtration wrong), or reaction in skin outside the marked field (the beam is bigger than you think, or scatter is leaking).
Note Troubleshooting. Output falls steadily week by week → the tube is ageing, the target is pitted, or the filter has shifted. Recalibrate; replace the tube. Output jumps about between days → an unstable high-voltage supply or mains variation. Fit a voltage stabiliser and measure the tube current during every treatment. The reaction is worse on one side of the field → the beam is off-centre in the applicator, or the applicator is not square on the skin. Skin reaction far heavier than expected for the dose → filter missing or wrong; stop and measure the half-value layer. The patient's position differs daily → marks fading, no shell, couch locks slipping. Tattoo the marks, make a shell, fix the couch. Electrometer readings drift with no beam → leakage over damp insulators. Dry everything, use a guarded cable, and keep the chamber warm.
Dangers
Danger An abandoned radioactive source is one of the deadliest objects left in the ruins of the old world. Teletherapy heads, brachytherapy needles, industrial radiography cameras and level gauges contain sealed capsules of cobalt-60, caesium-137 or iridium-192 that stay lethal for decades. They are small, heavy, and look like nothing: a stainless steel cylinder the size of a thumb, a lead pot, a wheeled machine head in an empty clinic. They will kill anyone who opens them.
This has happened repeatedly. At Goiânia, Brazil, in 1987, two scavengers took the rotating head from a caesium-137 therapy machine out of an abandoned clinic and broke it open; the glowing blue salt was shared around a neighbourhood as a curiosity, children handled it, and four people died, 20 were seriously ill, 249 were contaminated and whole houses had to be demolished. At Ciudad Juárez, Mexico, in 1983 a disused therapy source was sold for scrap, its 6,000 cobalt-60 pellets were melted into reinforcing bar and table legs and distributed across two countries. At Samut Prakan, Thailand, in 2000 a therapy head bought in a scrapyard was cut open with a hacksaw and three men died. At Mayapuri, Delhi, in 2010 a university irradiator sold for scrap killed one man and injured several.
The rule: if an unfamiliar heavy metal object — especially a cylinder, a pot or a machine head — carries a three-bladed symbol, or if it makes a counter or scintillation counter race, put it down, walk away, mark the spot, keep everyone at least 30 metres back, and fetch someone with instruments. Never open it, cut it, melt it, carry it in a pocket or take it home, and do not try to recover it "for science". The people who died of this were curious and uninformed, and they died badly, over weeks.
Also, in a working department:
- Never treat without a written prescription (total dose, dose per fraction, number of fractions, energy, filter, field size, distance, angles) and without two people checking the setup and the calculation.
- The spinal cord has no second chance. Keep a running total of the dose to it across every field and every course a patient has ever had. A patient returning years later for a second course near the same place must have the first course's doses retrieved — so the record must be permanent.
- Shield the room, not the operator. At 250 kV, a tenth-value layer is about 3 mm of lead or 9 cm of concrete; a wall the beam can point at wants five of those — roughly 15 mm of lead or 45 cm of concrete — and the others, which see only scatter and leakage, about half as much. The door opens into a bent passage, not into the beam line. Everyone working there wears a film badge read monthly, with limits of 20 mSv a year for workers, 1 mSv a year for the public (Detecting Radiation).
- Do not irradiate a pregnant woman unless her life depends on it, and keep young children away from the department. Protect the lens, the gonads and the thyroid with blocks whenever they are not the target.
- Radiation causes cancer as well as curing it. A second cancer in the treated field 10–30 years later is a real risk, which is why children are treated with radiation only when there is no alternative.
- Machine software is a safety system. The Therac-25 accidents of 1985–87, in which six patients received massive overdoses because software could silently put the machine in a wrong state with no hardware interlock to catch it, are the permanent lesson: every dangerous condition must be prevented by a physical interlock, not by a program (Programming, Robots and Automation).
What this unlocks
Radiotherapy is the point at which several industries meet: high voltage, high vacuum, Tungsten and Filament Wire, precision mechanics, dosimetry, pathology, imaging and record-keeping. Building it builds all of them. The same beams, measured the same way, sterilise medical supplies (Sterilising Instruments and Dressings) and test welds and castings (Welding, Metal Casting). The dosimetry is the same as radiation protection for Nuclear Energy (Overview) and for everyone working with X-Rays.
How it was first done
X-rays were discovered in November 1895 and used to treat disease within months: Leopold Freund in Vienna treated a hairy mole in November 1896, Victor Despeignes in Lyon had irradiated a stomach cancer earlier that year, and Emil Grubbé in Chicago claimed treatments in January 1896. The first documented cures of skin cancer were reported by Thor Stenbeck and Tage Sjögren in Stockholm in 1899. Radium, isolated by Marie and Pierre Curie in 1898, was placed directly against or into tumours from 1903 onwards.
The early decades were a mess of huge single doses, terrible burns and dead practitioners. The decisive work was done at the Institut du Radium in Paris in the 1920s and 30s, where Claudius Regaud showed on rams' testes that a dose spread over days sterilised the tissue without destroying the skin, and Henri Coutard applied the principle to patients, curing cancer of the larynx in 1922 with small daily doses over several weeks. Fractionation has governed the subject ever since.
Voltages climbed to 200 kV in the 1920s and to a few million volts in the 1930s. The first cobalt-60 teletherapy units began treating patients in 1951, in Saskatoon under Harold Johns and in London, Ontario. The first medical linear accelerator treated a patient at the Hammersmith Hospital in London in 1953 — 8 MV, built by the Medical Research Council and Metropolitan-Vickers — and by the 1980s linacs had displaced cobalt almost everywhere. The gray is named after Louis Harold Gray, the English physicist who put the measurement of radiation dose on a sound footing.