Part 9 · Health and Medicine
Pulse Oximeters and Patient Monitors
Machines that watch a sick person's oxygen, pulse, heartbeat and blood pressure without a pause, and call for help when something goes wrong. How a pulse oximeter reads oxygen with red and infrared light, how heart-rate and ECG monitors pick up the heart's electricity, how a cuff measures pressure by itself, and how to build simple versions and prove them against a careful nurse.
- Difficulty
- ●●●●●
- Time
- An evening to build a pulse beeper; two to four weeks for a working oximeter or ECG; months to prove a monitor against nurses' counts before trusting it on a ward
- Steps from zero
- 77 prerequisites →
This unlocks
—The full path from nothing — 77 steps before this one
- Fire
- Cordage and Rope
- Stone Tools
- Finding and Preparing Clay
- Finding and Identifying Ores
- Sulfur
- How the Body Works
- Writing and Numbers
- Hunting and Trapping
- Basic Woodworking
- The Kiln
- Pottery
- Charcoal
- Examining a Patient
- Arithmetic and Fractions
- Hides and Leather
- Lime
- Bricks
- Salt
- Mercury
- Lye and Potash
- Fermentation
- The Wheel and Axle
- Logic and Binary
- Glass
- Firebrick and Refractories
- Bellows
- Saltpeter
- Acids, Bases and Testing
- The Lathe
- Gears
- Glass Fibre and Glass Wool
- Copper Smelting
- Iron (the Bloomery)
- Lead
- Laboratory Glassware
- Sulfuric Acid
- Pumps
- The Waterwheel
- Lenses
- Blacksmithing
- Gold and Silver
- The Blast Furnace and Cast Iron
- Bronze
- Hydrochloric Acid
- Identifying Substances
- The Battery
- The Vacuum Pump
- Magnets
- Steel
- Chimneys, Stoves and Heating
- Selenium and Tellurium
- Bearings and Lubrication
- The Steam Engine
- The Light Bulb
- Copper Wire
- Electrolysis
- Photography
- Hydrogen, Oxygen and Gases
- Centrifugal Pumps and Fans
- Machine Tools
- The Generator
- The Vacuum Tube
- The Electric Arc Furnace
- Ammonia
- Ice and Refrigeration
- Pure Silicon
- Ventilation and Air Conditioning
- Semiconductors and the Transistor
- Clean Rooms
- Photocells and Sensors
- Integrated Circuits
- Making Microchips
- The Computer
- Programming
- Microprocessors and Personal Computers
- Microcontrollers
- Pulse Oximeters and Patient Monitors
Do them roughly in this order. See it on the tech tree →
Why it matters
A nurse counts a pulse and takes a blood pressure every hour (Examining a Patient, Nursing the Sick). The dangerous minutes come between the counts: an anaesthetised patient whose airway quietly blocks, a newborn who stops breathing. A monitor measures without stopping and sounds an alarm when a number crosses a limit. It does not replace the nurse; it calls the nurse.
Four machines do most of this work:
- the pulse oximeter, which reads how full of oxygen the blood is, through a fingertip, with no needle;
- the heart rate monitor, which counts every beat;
- the ECG monitor (electrocardiograph), which draws the heart's electrical rhythm;
- the automatic blood pressure monitor, which pumps a cuff and reads it by itself every few minutes.
The oximeter matters most. Lack of oxygen is invisible until the lips turn blue, at about 80 % saturation. An oximeter says when to start and stop Medical Oxygen and makes Anaesthesia and Breathing Machines and Ventilators far safer.
How it works
Light through a finger
Blood is red because of haemoglobin, the protein in red cells that carries oxygen. Haemoglobin loaded with oxygen and haemoglobin without it absorb light differently:
- Red light (wavelength 660 nanometres, deep red) is absorbed strongly by empty haemoglobin and only weakly by full haemoglobin. That is why venous blood looks dark and arterial blood bright.
- Infrared light (about 940 nm, invisible) is absorbed a little more by full haemoglobin than by empty.
Shine both colours through a fingertip onto a light sensor. Skin, bone and venous blood soak up most of the light steadily. With every heartbeat a little extra arterial blood swells into the finger and the light dips by about 1–2 %. That pulsing part (AC) belongs to arterial blood alone; dividing it by the steady part (DC) cancels out skin colour, finger size and lamp brightness.
The oximeter calculates the ratio of ratios:
R = (red pulse ÷ red steady) ÷ (infrared pulse ÷ infrared steady)
With full blood the red pulse is small, so R is low, about 0.4–0.5. As oxygen falls, R rises: about 1.0 at 80–85 % saturation. The relation was found by measuring volunteers, not by theory. A good working rule is SpO₂ ≈ 110 − 25 × R (as a percentage). SpO₂ means "saturation, measured by pulse oximetry".
The heart's electricity
Each heartbeat starts as an electrical wave that sweeps across the heart muscle. The body is salty water and conducts it, so a weak copy, about 1 millivolt, reaches the skin. An ECG picks this up between electrodes on the arms and leg, amplifies it about 1,000 times, and draws it as a line. The main hazard is the 50 or 60 Hz hum of mains wiring, which reaches the body from every cable in the room and can be a hundred times larger than the heart signal. The trick is a differential amplifier: it amplifies only the difference between two electrodes and ignores what both pick up equally.
Feeling the pulse in a cuff
As a cuff is let down slowly, each heartbeat makes its pressure wobble slightly. The wobbles are largest when the cuff pressure equals the mean arterial pressure, the artery's average pressure. From the rise and fall of the wobbles a machine finds the top (systolic) and bottom (diastolic) pressures, with no stethoscope: the oscillometric method.
What you need
| Item | For | From |
|---|---|---|
| Red (660 nm) and infrared (940 nm) LEDs, one each | Oximeter probe | Germanium, Gallium and Compound Semiconductors, Semiconductors and the Transistor |
| Silicon photodiode or phototransistor | Oximeter and pulse probe | Photocells and Sensors, Pure Silicon |
| Microcontroller with a 10- or 12-bit ADC | Every monitor here | Microcontrollers |
| Low-noise amplifiers (transistor or chip) | ECG, oximeter | Amplifiers and Oscillators, Integrated Circuits |
| Batteries, 6–9 V | Power, for safety | The Battery, Lithium Batteries |
| Silver wire or sheet, salt | ECG electrodes | Gold and Silver, Salt |
| Rubber bladder, tube, small pump, valve | Blood pressure monitor | Rubber, Bellows to Air Compressors |
| Pressure sensor (strain gauge on a diaphragm) | Blood pressure monitor | Photocells and Sensors, Pressure and Vacuum Gauges |
| Buzzer, display | Alarms and readings | Electric Bells, Buzzers and Alarms, Displays, Printers and Data Storage |
| A mercury blood pressure meter, a watch with seconds hand | Checking everything | Chronic Diseases, The Mechanical Clock |
Where to find it
Everything here is made, except silver (Gold and Silver) and salt. The hardest parts are the red and infrared LEDs, crystals of gallium arsenide and gallium arsenide phosphide (Germanium, Gallium and Compound Semiconductors). Before them, a lamp with coloured filters does the job less well (below).
The pulse oximeter
The finger probe
- Make the clip. Two jaws of hard wood, bone or moulded plastic, each about 45 mm long and 20 mm wide, hinged at the back and closed by a light spring (Springs). Line each jaw with a channel of dark rubber or felt that shuts out room light. Grip no harder than a clothes peg; a tight clip squeezes the blood out and kills the signal.
- Fit the lights. Set the red and infrared LEDs side by side in the upper jaw, about 3 mm apart, facing down, over the middle of the fingernail.
- Fit the sensor. Set the photodiode in the lower jaw directly opposite, under the pad of the fingertip. The light path through an adult finger is about 10–15 mm. Cover the sensor with a thin clear window so that it can be wiped clean.
- Cable. Five or six fine wires in a flexible sheath, 1–1.5 m long, shielded with a braid connected to the circuit's ground.
Good probe: blocks room light completely (the reading does not change when you turn a lamp on and off), fits fingers from a child's to a farmer's, and leaves no mark after an hour. Bad probe: light leaks round the sides, the jaws pinch, or the LEDs sit off-centre so the light skims past the finger.
The circuit
- Take turns. The microcontroller switches the LEDs in a repeating cycle of four steps: red on, dark, infrared on, dark. Each step lasts about 250 microseconds, so each colour is read about 1,000 times a second. Drive each LED at 10–30 mA through a transistor and a resistor; adjust the current so the steady signal sits at about half the ADC's range.
- Subtract the dark. The "dark" readings measure room light. Subtract each from the colour reading before it. This removes daylight and lamp flicker.
- Amplify. The photodiode gives a tiny current. Turn it into a voltage with a high-value resistor (about 1 megohm) in the feedback loop of an amplifier, so a few microamps give a few volts.
- Separate pulse from steady. In the program, average each colour over 2–3 seconds to get the steady part (DC). Track each pulse's highest and lowest point to get the pulsing part (AC). Average the result over 4–8 beats to smooth out movement.
- Calculate R, then SpO₂, from the table below. Count the time between pulse peaks for the heart rate: rate = 60 ÷ seconds between beats.
- Display the numbers and, if you can, the pulse wave: a clean, regular wave is the best sign the reading is true.
| R | SpO₂ (approx.) |
|---|---|
| 0.4 | 100 % |
| 0.6 | 95 % |
| 0.8 | 90 % |
| 1.0 | 85 % |
| 1.2 | 80 % |
| 1.4 | 75 % |
Calibration. These numbers are a starting point. The real curve depends on your exact LED colours and your probe, and makers find it by testing volunteers breathing thin air while their arterial blood is sampled. You cannot do that safely. Do this instead:
- Healthy people at rest at low altitude read 95–99 %. Test twenty; if they read 90 %, shift the table.
- Holding the breath for 30–40 seconds should make the reading dip a few points 20–30 seconds later, then recover.
- With even one factory-made oximeter, put both probes on neighbouring fingers of many patients and adjust your table until they agree within 2 points above 90 %.
- Trust a home-made oximeter for trends first. Trust its exact number below about 85 % only after long comparison.
Before LEDs
The first oximeters, in the 1930s and 1940s, shone a small torch bulb through the ear lobe onto two Selenium and Tellurium photocells behind red and infrared glass filters. You can do the same with a 2–3 W bulb and the same arithmetic. It is warm, power-hungry and less accurate, but it shows a falling saturation in a ward with nothing else.
The heart rate monitor
The simplest useful monitor of all is a pulse beeper: one LED (infrared or green), one photosensor in a finger clip or an ear clip, an amplifier, and a buzzer that chirps once per beat. Build the probe exactly as above with one light. The program finds each pulse peak, works out the rate, and sounds an alarm when it leaves the limits.
A beep per beat lets an anaesthetist keep their eyes on the patient and hear the pulse slow. Make the pitch fall as the saturation falls, as modern machines do, and a trained ear hears trouble before any number is read. The rate can also come from the ECG's tall spikes, which still works when cold fingers or shock defeat a finger probe.
Good monitor: agrees with a hand count within ±5 beats a minute, and does not double-count (each tall wave counted twice) or miss beats when the patient moves a little.
The ECG monitor
The electrode
An ECG electrode must touch the skin through salty paste and must not make its own drifting voltage. Bare metal on skin acts like a weak battery that wanders by tens of millivolts and swamps the heart. Silver coated with silver chloride does not.
- Cut discs of silver sheet 10–15 mm across, or coil 5 cm of silver wire flat. Solder or crimp a fine wire to each.
- Chloride them: hang the silver in strong salt water (a heaped tablespoon of salt in a cup of water) and connect it to the positive pole of a 1.5 V cell through a 100 ohm resistor, with any other piece of silver or carbon on the negative pole. After 10–20 minutes the silver turns an even dark grey-brown to purple. Rinse.
- Set each disc in a shallow cup of rubber or plastic 25–35 mm across, which holds a sponge or a dab of paste and is held on with a strap or sticky plaster.
- Paste: salt water thickened with starch or Agar and Culture Plates, about 5 % salt, so it stays put and keeps the skin moist.
Einthoven's patients simply put their hands and a foot into buckets of salt water with metal plates in them, which works for a patient who can sit still.
Where the electrodes go
Use three electrodes, on the right arm (RA), left arm (LA) and left leg (LL) — on the wrists and ankle, or close to the shoulders and on the lower left belly for a lying patient, which reduces muscle noise. A fourth on the right leg connects to the circuit's ground or a "driven" ground, which cancels hum. Comparing pairs gives the three standard leads (views):
- Lead I = left arm minus right arm
- Lead II = left leg minus right arm (the usual monitoring lead: it shows the clearest waves)
- Lead III = left leg minus left arm
The amplifier and display
- Input stage: a differential amplifier with very high input resistance (over 10 megohms), so the electrodes are not loaded, and with good common-mode rejection (ignoring what both inputs share). Give it a gain of about 10.
- Filter: let through only 0.5–40 Hz for monitoring (0.05–100 Hz for a full diagnostic recording). This removes slow drift from breathing and the fast fuzz of muscle.
- Second stage: a further gain of about 100, so 1 mV of heart signal becomes about 1 V.
- Display: a cathode-ray tube sweeping left to right, a pen on a moving strip of paper (Instruments, Chart Recorders and Control Rooms), or the microcontroller sampling 250–500 times a second and drawing on a screen.
- Standard scale: paper moving at 25 mm a second and 10 mm for 1 mV. Then a small 1 mm square is 0.04 s and a big 5 mm square is 0.2 s. Build in a button that injects exactly 1 mV, so the operator sees a 10 mm step and knows the scale is right.
Reading it, in brief. The P wave is the upper chambers firing; the tall QRS is the main chambers; the T wave is them recovering. Normal: P to start of QRS 0.12–0.20 s (three to five small squares); QRS narrower than 0.12 s. Heart rate = 300 ÷ the number of big squares between two R spikes. Above all the monitor must show at once the rhythms that kill: a shapeless wiggle (ventricular fibrillation) or a flat line (asystole, but first check the electrodes). Finer reading takes training.
The automatic blood pressure monitor
Build the cuff as in Chronic Diseases and Medical Instruments, with the right size bag for the arm. Then add:
- A pump: a small diaphragm pump driven by a battery motor (The Electric Motor), able to reach 300 mmHg in a cuff in under 15 seconds.
- Two valves: an electrically worked step valve that lets out a few mmHg at a time, and a dump valve that empties the cuff at once.
- A pressure sensor: a strain gauge bonded to a thin diaphragm (Photocells and Sensors), read by the ADC. Calibrate it against a mercury column at 0, 100, 200 and 250 mmHg.
- The program:
- Pump to 160–180 mmHg for the first reading, or 30 mmHg above the last systolic after that.
- Let out 5–8 mmHg at a time. At each step, wait for two heartbeats and record the size of the wobble.
- The step with the largest wobble is the mean arterial pressure (MAP).
- Systolic is the pressure, above the MAP, where the wobble is about 50–55 % of the largest.
- Diastolic is the pressure, below the MAP, where the wobble has fallen to about 75–80 % of the largest.
- Dump the cuff. Repeat every 5, 15 or 60 minutes as set.
These fractions differ a little between machines and must be adjusted by comparing with careful hand readings (below). As a cross-check, MAP should be close to diastolic + one-third of the difference.
Safety built in: a mechanical relief valve that opens at 300 mmHg whatever the program does; a timer that dumps the cuff if any reading takes longer than 2 minutes; and never more often than every 2–3 minutes for long, or the arm swells and bruises.
Alarms
A monitor that cannot call for help is only a display. Give every monitor:
| Alarm | Usual adult limits |
|---|---|
| Low SpO₂ | below 90 % (or 88 % in long-standing lung disease) |
| Heart rate | below 50 or above 120–130 a minute |
| Blood pressure | systolic below 90 or above 180 mmHg |
| No pulse found, probe off, electrode off | always on |
| Battery low | always on |
Children have higher normal rates; set limits for each patient. Use one sound for "danger to the patient" (urgent, high) and another for "machine fault" (slower, lower). A silence button quiets an alarm for at most 2 minutes. Too many false alarms teach nurses to ignore them (alarm fatigue): cut them by averaging over a few seconds and fixing loose probes, never by turning alarms off.
How to tell it worked
Check every monitor against a nurse's count. Do it on many patients, not one, and write the results down (Statistics and Probability).
| Test | Pass |
|---|---|
| Heart rate vs a full 60-second hand count at the wrist | Within ±5 a minute on 20 patients |
| SpO₂ on 20 healthy adults at rest | All 95–99 % (at low altitude) |
| SpO₂ vs a factory oximeter | Within ±2 points above 90 %, ±4 below |
| ECG calibration pulse | 10 mm tall; small squares 0.04 s |
| ECG rate vs pulse count | Match within ±5, unless the pulse is irregular |
| Automatic BP vs a careful nurse with mercury meter and stethoscope, same arm, readings alternating | On average within 5 mmHg, and three-quarters within 10 mmHg |
| Alarms | Each one sounds when its limit is crossed, tested by unplugging the probe and by setting a limit just above the patient's value |
Note Common faults. Oximeter shows nothing or jumps about → cold fingers (warm the hand), movement, a tight clip, or room light leaking in; try another finger or the ear lobe. Reads 100 % on someone who is clearly ill → carbon monoxide poisoning, see Dangers. Reads low on everyone → nail varnish or dirt on the nail, or the table needs shifting. ECG thick and fuzzy → mains hum: check the right-leg electrode, keep leads close together and away from cables, run on battery. ECG wanders up and down → dry paste or bare metal electrodes; re-chloride. Blood pressure fails or reads high → wrong cuff size, a leak, or the patient moving or shivering.
Dangers
Danger Electric shock through the heart. An ECG deliberately connects wires to a patient's body. A fault in a mains-powered machine can drive current straight through the chest. Run patient-connected circuits from batteries only, and if they must link to a mains-powered display or computer, pass the signal through a light link (an LED and a photocell in a sealed tube) so no wire connects the two. Take special care with a patient who has a pacing wire or a fluid-filled tube leading into the heart: along that path a current far too small to feel can stop it, so use only a battery-powered, isolated monitor, and never touch the wire and the machine at once.
Warning The machine can lie. Treat the patient, not the number.
- Carbon monoxide (smoke, charcoal stoves, engine fumes) makes haemoglobin look full of oxygen. A poisoned person can read 98 % while dying. Suspect it whenever several people in one room are ill at once.
- On dark skin many oximeters read 2–4 points too high at low saturations. Give oxygen earlier, not later.
- Shock, cold and movement give a weak, false reading. If the wave on the screen is not a clean pulse, do not believe the number.
- A flat ECG line is usually a loose electrode. Look at the patient and feel the pulse before you start CPR (Emergencies), and never delay CPR to fiddle with wires.
- Pressure injury: a finger clip left on one finger for many hours, especially on a baby, can burn or bruise the skin. Move it every 2–4 hours. A blood pressure cuff cycling too often damages the arm.
- Over-trust: a monitored patient still needs to be looked at and spoken to every hour.
What this unlocks
Monitors make the risky machines possible: Breathing Machines and Ventilators, Kidney Dialysis, long operations (Basic Surgery) and intensive care in a hospital. More colours of light lead to Analytical Instruments for blood; ECGs sent by wire or radio (The Telephone, Telemetry and Remote Control, The Internet) let a distant expert read them.
How it was first done
Augustus Waller in London recorded the first human heart trace in 1887 with a capillary electrometer. Willem Einthoven in Leiden built the far more sensitive string galvanometer in 1901–1903, a silvered quartz thread between magnet poles, its shadow photographed on moving film. He named the P, Q, R, S and T waves, defined the three limb leads, and won the Nobel Prize in 1924.
Scipione Riva-Rocci introduced the mercury cuff in 1896, and Nikolai Korotkov the listening method in 1905. Étienne-Jules Marey had noticed in 1876 that the pulsations in a pressure chamber were largest at a middle pressure, the basis of the oscillometric method; automatic machines using it spread in hospitals from the late 1970s.
Karl Matthes in Germany built a two-colour ear oximeter in 1935, and Glenn Millikan a practical one for pilots in the Second World War. In 1972–74 Takuo Aoyagi of Nihon Kohden in Tokyo realised that the pulsing part of the light belonged to arterial blood alone: the pulse oximeter. Finger-probe oximeters became standard in operating theatres in the 1980s.
Source: content/entries/health/patient-monitors.md