Part 13 · Electricity
Measuring Electricity
Volts, amps, ohms and watts in plain words, and Ohm's law that ties them together. How to set up your own standards from a copper bath, a Daniell cell and a length of wire, and how to turn the galvanometer into real ammeters, voltmeters, resistance boxes and a bridge that measures resistance to better than one part in a hundred.
- Difficulty
- ●●●●●
- Time
- A day to set up the standards; a week to build a moving-coil meter, shunts and a bridge
- Steps from zero
- 18 prerequisites →
You need first
The ElectromagnetThis unlocks
Electricity Meters and BillingThermocouples and ThermoelectricityInstruments, Chart Recorders and Control RoomsResistors, Capacitors and CoilsElectronic Test EquipmentThe full path from nothing — 18 steps before this one
- Fire
- Cordage and Rope
- Stone Tools
- Finding and Preparing Clay
- Finding and Identifying Ores
- Hunting and Trapping
- Basic Woodworking
- Charcoal
- Pottery
- Hides and Leather
- Fermentation
- Bellows
- Copper Smelting
- Iron (the Bloomery)
- Blacksmithing
- The Battery
- Copper Wire
- The Electromagnet
- Measuring Electricity
Do them roughly in this order. See it on the tech tree →
Why it matters
The galvanometer in The Electromagnet tells you whether a current flows and whether it is bigger than before. That finds a broken wire; it does not let you design anything. To choose a wire's thickness, the number of cells for a lamp, whether a The Generator gives what it should, how long a The Rechargeable Battery will last, or how much copper an Electrolysis bath will deposit, you need numbers that mean the same tomorrow, in another workshop, and in this book.
This chapter gives you the four quantities in plain words, Ohm's law, three standards you can make yourself, and proper instruments: the moving-coil meter, ammeters, voltmeters, a resistance box and a Wheatstone bridge. Every later electrical chapter — Wiring and Electrical Safety, The Transformer, The Telegraph, Power Lines and Distribution, Radio — assumes you can measure.
How it works
The four quantities
Think of electricity in a wire as water in a pipe.
| Quantity | Unit | Symbol | Water picture | Example |
|---|---|---|---|---|
| Voltage — the push | volt (V) | V | Pressure, or the height the water falls | A Daniell cell: 1.1 V. A lead-acid cell: 2 V |
| Current — the flow | ampere, amp (A) | I | Litres per second through the pipe | A small electromagnet: 0.5 A. A carbon lamp: 1 A |
| Resistance — how hard it is to push through | ohm (Ω) | R | How narrow and long the pipe is | 1 m of 1 mm copper wire: 0.022 Ω |
| Power — the rate of doing work | watt (W) | P | The work the falling water can do each second | A small dynamo: 100–500 W |
A milliamp (mA) is a thousandth of an amp; a kilowatt (kW) a thousand watts. Energy is counted in kilowatt-hours (kWh): a 100 W lamp burning 10 hours uses 1 kWh. A horse working steadily gives about 750 W.
Ohm's law
In 1827 Georg Ohm found that, for a wire at a steady temperature, the current is proportional to the voltage: double the push, double the flow. The ratio is the resistance:
V = I × R — volts equal amps times ohms. I = V ÷ R and R = V ÷ I
Examples:
- A coil of 2.2 Ω across a Daniell cell of 1.1 V: I = 1.1 ÷ 2.2 = 0.5 A.
- A lamp that takes 0.5 A at 50 V has a resistance (hot) of 50 ÷ 0.5 = 100 Ω.
Power and heat
P = V × I — watts equal volts times amps. Because V = I × R, also P = I² × R.
The heat in a wire goes with the square of the current: double the current and it heats four times as fast. That is why power travels long distances at high voltage and low current (The Transformer, Power Lines and Distribution). Heating water makes an honest power meter: a coil in 1 litre of water that warms it 1 °C per minute is using about 70 W (The Thermometer).
Resistances together
- In series (one after another): add them. 2 Ω + 3 Ω = 5 Ω.
- In parallel (side by side): the total is less than the smallest. Two equal resistances in parallel give half of one. In general, 1/R = 1/R₁ + 1/R₂ + …
Every cell has its own resistance
The liquid and plates inside a cell have their own internal resistance, so the terminal voltage falls as the current rises: terminal volts = open-circuit volts − (current × internal resistance). A Daniell cell has roughly 0.5–2 Ω inside; a large lead-acid cell about 0.01 Ω. That is why a Daniell cell never gives more than an amp or two, while a lead-acid cell can melt a thick wire.
What resistance depends on
A wire's resistance goes up with its length, down with its cross-section (area), and depends on the metal:
R = (resistivity × length in metres) ÷ (cross-section in mm²)
| Metal | Resistivity (Ω for 1 m of 1 mm² wire) | Change per °C | Use |
|---|---|---|---|
| Copper, pure, annealed | 0.017 | +0.4% | Connections, coils |
| Brass | 0.06–0.08 | +0.15% | Terminals, springs |
| Iron, soft | 0.10–0.14 | +0.6% | Long lines, cheap resistances |
| German silver (copper 60, nickel 20, zinc 20) | about 0.3 | +0.03% | Standard resistances, shunts |
The cross-section of a round wire is 0.785 × diameter² (in mm). For 1 mm wire that is 0.785 mm²; for 0.5 mm, 0.196 mm². Measure the diameter by winding 20 turns side by side, as in Copper Wire.
Copper and iron change a lot as they warm — a copper coil is 10% more resistant at 45 °C than at 20 °C. German silver hardly changes, so use it for anything that must keep a fixed value. It needs nickel (Nickel and Cobalt); melt the copper under charcoal, add the nickel, then the zinc last (Zinc and Brass).
What you need
| Item | Notes |
|---|---|
| A galvanometer | From The Electromagnet; you will build a better one below |
| Copper sulfate, sulfuric acid, alcohol | For the copper coulometer (The Battery, Sulfuric Acid, Distillation) |
| A balance weighing to 0.01 g | Measurement from Scratch. Essential — it is your link to the ampere |
| Two or three good Daniell cells | The Battery. One kept only as a standard |
| Pure copper wire, 1 mm | 50 m, for a resistance standard (Copper Wire) |
| Iron or German silver wire | 0.3–0.5 mm, a few hundred metres, for resistance coils |
| Hardened steel magnets | Two or three horseshoes about 10 cm long, for the moving-coil meter (Magnets) |
| Thin brass or bronze strip | For hairsprings (Springs, Zinc and Brass) |
| Silk thread, shellac, wax | Covering and varnishing coils (Copper Wire) |
| A thermometer | For noting the temperature of standards (The Thermometer) |
| Dry hardwood, brass screws, card | Bases, terminals, scales (Basic Woodworking) |
Where to find it
Everything here comes from earlier chapters (see the table). Nickel for German silver comes from green garnierite ore (Nickel and Cobalt); without it, use soft iron wire for resistances and keep it at a steady temperature.
Step 1: Make your own standards
The ampere: the copper coulometer
Current through copper sulfate lays down copper on the negative plate at an exact rate: 1.186 g for every ampere flowing for one hour (Electrolysis). Weigh the copper and you have measured the current. This is a coulometer (or voltameter).
- The bath. In a glass or glazed jar of about 1 litre, dissolve 150 g of copper sulfate crystals in 1 litre of rainwater, then stir in 50 g (about 27 ml) of strong sulfuric acid — acid into water, slowly — and 50 ml of alcohol. The acid and alcohol make the copper lay down smooth and firm.
- The plates. Three strips of pure copper sheet, each about 10 × 5 cm: two outer ones joined together as the positive (anode), and the middle one, the negative (cathode), hanging between them, 2–3 cm from each.
- Size the cathode so that the current is between 0.2 and 2 amps for each 100 cm² of its surface (counting both faces). Too much current makes spongy copper that falls off; too little is merely slow.
- Clean the cathode with fine sand, rinse it, dip it in the bath without current for a minute, rinse in clean water, then alcohol, dry it in warm air and weigh it.
- Connect the coulometer in series with your circuit and the meter you want to calibrate. Run for an exact time — at least 30 minutes, timed with a clock or a The Mechanical Clock — keeping the current steady.
- Rinse, dry and weigh the cathode again.
Current (A) = gain in grams ÷ (1.186 × hours)
Example: the cathode gained 0.89 g in 45 minutes (0.75 h): 0.89 ÷ (1.186 × 0.75) = 1.00 A.
The volt: the Daniell cell
A Daniell cell gives 1.10 V (1.07–1.11) whatever its size, so long as the copper sulfate is saturated and the zinc clean. Build one with care (The Battery) and use it only as a standard: never draw current from it except for a moment through a high resistance, or in the potentiometer below.
The ohm: a length of wire
45.6 m of pure, annealed copper wire exactly 1 mm across has a resistance of 1 Ω at 20 °C. Wind it bifilar — doubled back on itself at its middle and both halves wound together — so the current in neighbouring turns runs opposite ways and the coil makes no magnetism. Use it to adjust a 1 Ω coil of German silver (about 0.65 m of 0.5 mm wire), which will not change with the weather.
Check them against each other
Since R = V ÷ I, the three must agree. Drive current from your standard cell through a known resistance, measure it with the coulometer, and check Ohm's law. More than about 3% out means impure copper, a wrong diameter, or a stale cell.
Step 2: Calibrate the galvanometer you have
Put your galvanometer (best, the tangent galvanometer of The Electromagnet) in series with the coulometer, a battery and a long iron wire with a sliding clip. At five or six settings, weigh the copper and note the needle's angle; draw the curve on squared paper (Arithmetic and Fractions). For a tangent galvanometer one reading is enough — if 45° means 0.30 A, 26.6° means 0.15 A — but it must stay where it was calibrated, since the Earth's magnetism and nearby iron affect it.
Step 3: The moving-coil meter
The best meter of the 19th century, still in use today: a light coil turns between the poles of a strong permanent magnet, held back by a hairspring. Its divisions are evenly spaced, it does not care which way it faces or what iron is nearby, and it can be carried about.
- The magnet. Two or three hardened steel horseshoe magnets about 10 cm long, clamped side by side with like poles together, magnetised as strongly as you can (Magnets).
- Pole pieces. Soft iron blocks on the magnet's ends, their facing sides hollowed into a round bore about 3 cm across.
- The core. A soft iron cylinder 2.6 cm across, held in the middle of the bore by a brass bracket, leaving an even 2 mm gap all round for the coil.
- The coil. A rectangular frame of thin copper or brass foil, 2.8 cm wide (so that it fits round the core, its two sides running in the gap) and 2.5 cm tall, wound with 100 turns of 0.2 mm silk-covered wire (about 5–6 Ω) and shellacked. Currents induced in the metal frame as it swings brake it, so the pointer settles without bouncing (damping).
- Pivots. Sharp hardened steel points at top and bottom of the frame, turning in conical hollows in hard brass or polished glass beads held in brass bridges. The coil must swing freely without touching anything.
- Hairsprings. Two flat spirals of thin hard brass or bronze strip (about 0.1 × 0.5 mm, 5–6 turns), one at each pivot, wound opposite ways so warmth affects them equally. They pull the coil back to zero and carry the current in and out.
- Pointer and scale. A light straw or thin brass pointer 10 cm long, counterweighted, over a card scale of about 90°.
- Calibrate with the coulometer at full and half scale; a well-centred coil gives evenly spaced marks.
A meter like this typically swings full scale for 5–20 mA. That alone makes it a milliammeter. The next two steps turn it into anything else.
Tip More sensitive still: the mirror. Hang the coil from a fine bronze or silver wire instead of pivots, glue a tiny mirror to it, and shine a lamp's beam off the mirror onto a scale 1 m away. The spot of light is a weightless pointer 2 m long. Such a galvanometer detects a millionth of an amp and was used to receive messages on undersea The Telegraph cables.
Step 4: Ammeters and voltmeters
The ammeter: add a shunt
To measure large currents, let most of the current bypass the meter through a shunt — a low resistance connected side by side with it. The meter then carries a fixed small fraction.
Shunt resistance = meter resistance × meter current ÷ (full current − meter current)
For a meter of 5 Ω that reads full scale at 10 mA, to read 1 A: 5 × 0.01 ÷ 0.99 = 0.0505 Ω. For 10 A, 0.005 Ω.
Make the shunt from a German silver (or iron) strip thick enough not to warm up, between heavy copper blocks; make it a little long, then shorten it until the meter agrees with the coulometer. A shunt belongs to one meter. An ammeter always goes into the line, in series with what you are measuring.
The voltmeter: add a series resistance
To measure voltage, put a high resistance in series with the meter. The current through it is then proportional to the voltage across its two ends.
Series resistance = full-scale volts ÷ meter current − meter resistance
For the same 10 mA meter to read 10 V full scale: 10 ÷ 0.01 − 5 = 995 Ω — about 260 m of 0.2 mm iron wire, or 105 m of 0.2 mm German silver, wound bifilar on a bobbin. Check it on your Daniell cells: one should read 1.1 V. A voltmeter always goes across the two points you compare. It draws a little current, which pulls the reading down; aim for at least 100 Ω per volt of full scale.
The moving-iron meter: simple, rugged, and works on AC
The moving-coil meter only reads direct current: alternating current swings it back and forth too fast to follow, and it reads zero. For a The Transformer or an AC The Generator, build a moving-iron meter.
Wind a short coil on a tube about 3 cm across — 10–30 turns of thick wire for amps, 1,000 or more of fine wire for volts — and pivot a thin curved vane of soft iron so it can swing into the coil's mouth, with a pointer and a small weight or hairspring to pull it back. Current either way pulls the iron in, so it reads AC. Mark the crowded scale point by point against known DC currents.
Step 5: Measuring resistance
The quick way
Measure the current through it and the voltage across it, and divide: R = V ÷ I. Accurate to a few per cent.
The resistance box
Wind German silver coils of 1, 2, 2, 5, 10, 20, 20, 50 and 100 Ω, each bifilar and adjusted against your standard, under a wooden lid, connected in a row between thick brass blocks on top. A tapered brass plug in each gap shorts its coil; pull the plug and the coil is added. Nine coils give any whole number of ohms from 1 to 210.
The Wheatstone bridge
The bridge compares an unknown resistance to a known one by balancing, so the galvanometer only has to show zero — it need not be calibrated at all.
- On a board 1.2 m long, stretch 1 m of bare German silver or iron wire, 0.5 mm, tight between two heavy copper strips, over a metre scale marked in millimetres.
- Leave two gaps in a thick copper strip above it: put your known resistance R (the box) in the left gap and the unknown X in the right gap.
- Connect a galvanometer from the junction between R and X to a sliding contact — a knife-edge of brass on an insulating handle — that you press onto the bare wire.
- Connect one or two cells across the two ends of the wire, through a press-button key so that the current flows only while you measure (warm wires change value).
- Press the key and touch the contact to the wire at various places. Find the point where the galvanometer does not swing at all. Read the lengths: a on the side of R, b on the side of X.
- X = R × b ÷ a. Choose R so that the balance falls near the middle of the wire; that is where the bridge is most accurate — better than 1%.
Comparing voltages exactly: the potentiometer
The same slide wire measures voltage without drawing current. Keep a steady current flowing along the wire from a lead-acid cell. Connect your standard Daniell cell, through a galvanometer, from one end of the wire to the sliding contact (positive to the same end as the lead-acid cell's positive) and find the balance length L₁; do the same with the unknown cell for L₂. Its voltage is 1.10 × L₂ ÷ L₁.
How to tell it worked
- Three ways agree. Measure one coil by volts ÷ amps, by the bridge, and from its length and diameter. All three should agree within 2–3%.
- Two meters in series read the same current. Two voltmeters across the same cell read the same voltage.
- Ohm's law test. Connect 1, 2, 3 and 4 Daniell cells in turn through a coil of about 20 Ω. The current should rise in nearly equal steps: about 0.05, 0.1, 0.15, 0.2 A.
- Reversal. A moving-coil meter reverses when the current reverses; a moving-iron meter reads the same either way.
- Repeat tomorrow. The coulometer, run twice at the same setting, gives the same gain within 1%.
Note Troubleshooting. Coulometer copper dark, spongy or flaking → current too high for the plate area; use a bigger cathode or less current. Coulometer result too low → copper knocked off while rinsing, or the cathode was not dry when weighed. Dry it in warm air for 15 minutes. Moving-coil pointer sticks → coil rubbing in the gap, or pivots too tight; check by blowing gently on the pointer. It should swing freely and settle. Pointer does not return to zero → hairspring overstrained or touching itself. Remake it with fewer turns. Bridge never balances → R and X swapped compared with the lengths you are reading, a bad contact at a strip, or the galvanometer connected to the wrong junction. Readings drift as you work → coils warming. Use the key, and German silver for anything that must not change.
Dangers
Danger Never put an ammeter across a battery or generator. An ammeter with its shunt is nearly a dead short. Across a lead-acid battery or a dynamo it can draw hundreds of amps, melt the shunt, burn the meter and throw molten metal. An ammeter always goes in series with something that limits the current.
- High voltage. Measuring a The Generator or a line of 50–100 V means touching live terminals. Connect meters with the machine stopped, use well-insulated leads, and keep one hand in your pocket (see Wiring and Electrical Safety).
- Acid. The coulometer contains sulfuric acid and copper sulfate: eye protection, wash splashes at once (Sulfuric Acid, The Battery).
- Hot shunts and resistances. Wires carrying more than they should glow and set wood alight. Mount power resistances on clay or stone, away from wood.
- Mercury (if you make a mercury standard or contacts) gives off poisonous vapour; see Mercury.
What this unlocks
- Proper design of coils, lines and lamps — Wiring and Electrical Safety, The Telegraph, The Light Bulb.
- Running and charging batteries by the numbers — The Rechargeable Battery, Electrolysis, Electroplating.
- Rating The Electric Motors and The Generators in watts, and selling power by the kilowatt-hour (Electricity Meters and Billing) on an Power Lines and Distribution.
- AC measurement for the The Transformer, and fine measurement for Radio and The Vacuum Tube work (Electronic Test Equipment).
- Science: careful measurement is the heart of Forces, Energy and Heat and the The Scientific Method.
How it was first done
In 1820 Johann Schweigger built the first galvanometer. Georg Simon Ohm, a schoolteacher in Cologne, published his law in 1827, having used thermocouples as a steady source and wires of different lengths and thicknesses; he was honoured only after the Royal Society of London gave him its Copley Medal in 1841. Claude Pouillet described the tangent galvanometer in 1837, and Michael Faraday found the laws of electrolysis in 1833–34. Samuel Hunter Christie invented the bridge in 1833 and Charles Wheatstone made it famous in 1843. William Thomson (later Lord Kelvin) built the mirror galvanometer for the Atlantic telegraph cable of 1858.
Units came later. Werner Siemens proposed a resistance standard in 1860 — a column of mercury 1 m long and 1 mm² in section. A committee of the British Association, from 1861 onwards, defined the ohm and the volt, and the International Electrical Congress in Paris in 1881 adopted the ohm, volt, ampere and coulomb. Jacques-Arsène d'Arsonval made the moving-coil galvanometer in 1882, and Edward Weston turned it into the rugged pointer meter with spiral springs in 1888. The units are named after Volta, Ampère, Ohm and James Watt.
Source: content/entries/electricity/measuring-electricity.md