Part 15 · Knowledge and Science
The Electron Microscope
Light cannot show anything smaller than about a fifth of a thousandth of a millimetre, but a beam of electrons in a vacuum can show things a thousand times smaller — viruses, the inside of cells, the grains and flaws in metals, the patterns on a microchip. Why electrons see finer detail, how a coil of wire in an iron case focuses them, how the transmission and scanning microscopes are built and how samples are made thin, dry and conducting for them — and a ladder of stepping stones from a cathode-ray tube to a working instrument.
- Difficulty
- ●●●●●
- Time
- A month for a two-lens electron-optical bench; one to three years and a team for a working transmission or scanning microscope
- Steps from zero
- 47 prerequisites →
This unlocks
—The full path from nothing — 47 steps before this one
- Fire
- Stone Tools
- Cordage and Rope
- Finding and Preparing Clay
- Finding and Identifying Ores
- Sulfur
- Basic Woodworking
- Hunting and Trapping
- The Kiln
- Pottery
- Charcoal
- Hides and Leather
- Lime
- Mercury
- Lye and Potash
- Fermentation
- The Wheel and Axle
- Glass
- Bellows
- Saltpeter
- The Lathe
- Copper Smelting
- Lead
- Iron (the Bloomery)
- Laboratory Glassware
- Sulfuric Acid
- Pumps
- Lenses
- Gold and Silver
- Blacksmithing
- Nitric Acid
- The Vacuum Pump
- The Battery
- The Microscope
- Silver Nitrate and Silver Salts
- The Light Bulb
- Copper Wire
- The Electromagnet
- The Vacuum Tube
- Mirrors
- The Induction Coil
- The Telegraph
- Vacuum Coating and Thin Films
- High-Vacuum Pumps and Gauges
- The Telephone
- Radio
- Cathode Rays and Television
- The Electron Microscope
Do them roughly in this order. See it on the tech tree →
Why it matters
The best The Microscope made of glass cannot show detail finer than about 0.2 µm (0.0002 mm), half the wavelength of light. That shows bacteria as shapes but nothing of the viruses that cause smallpox, measles, polio and influenza (Infectious Diseases), which are 20–300 nanometres across (a nanometre, nm, is a millionth of a millimetre). It shows nothing of the fine structure inside cells, of the tiny particles in paints and catalysts, of the grain boundaries and cracks where metals fail (Strength of Materials), or of the lines on a microchip.
An electron microscope magnifies 10,000 to 1,000,000 times. It identified the viruses behind diseases and became the everyday tool of metallurgy and chip-making.
Be honest about the scale of the job. It needs High-Vacuum Pumps and Gauges, Cathode Rays and Television tubes, steady high-voltage supplies and very fine machining; the first instruments took small teams several years. But the route is a ladder of instruments, each useful in itself, and the bottom rung is within reach of anyone who has built an oscilloscope.
How it works
Why electrons see smaller things
Electrons behave as waves as well as particles. The faster they go, the shorter their wavelength. For an electron speeded through V volts:
wavelength in nm ≈ 1.23 ÷ √V
| Accelerating voltage | Wavelength | Compared with green light (550 nm) |
|---|---|---|
| 100 V | 0.12 nm | 4,500 times shorter |
| 10,000 V | 0.012 nm | 45,000 times shorter |
| 50,000 V | 0.0055 nm | 100,000 times shorter |
The wavelength never limits a home-built instrument. Poor lenses, unsteady voltages, vibration and stray magnetic fields do. The first microscopes resolved about 50 nm, those of 1940 about 10 nm, the best today less than 0.1 nm. For a first instrument, aim for 20–50 nm — still five to ten times finer than light.
Lenses made of magnetic field
An electron passing through a magnetic field is pushed sideways, at right angles both to the field and to its own path. A short coil of wire has a field that bulges outward from its axis. An electron that enters it off-centre is twisted round the axis and, at the same time, pushed inward towards it: the field acts like a convex glass lens. Electrons that start from one point are brought together again at another point, and an image forms (Lenses).
- Strength. The focal length shortens as the square of the ampere-turns (turns of wire × current). Double the current and the focal length falls to a quarter. So the focus is adjusted by a knob on a current, not by moving the lens.
- Faster electrons are stiffer. Focal length grows in proportion to the voltage. A lens that focuses 10 kV electrons at 1 cm focuses 50 kV electrons at 5 cm.
- Iron makes it strong. Wrapping the coil in soft iron and leaving only a narrow gap at the bore — the pole pieces — squeezes the whole field into a few millimetres. That shortens the focal length to 1–3 mm, which is what an objective lens needs.
- The image rotates. Because the electrons spiral, the image turns as well as enlarges. It does not matter.
- Aberrations. Every magnetic lens bends the outer rays too strongly (spherical aberration). No convex-concave pair cancels it as in glass. The cure is to use only rays very close to the axis — a hole of 20–50 µm in a platinum disc, the objective aperture, set just behind the specimen. Electrons of slightly different speed focus at different places (chromatic aberration), so the high voltage and lens currents must be held steady to about one part in ten thousand for 20 nm work, better for finer. Lenses that are not perfectly round give a blurred, streaky image (astigmatism), corrected by a stigmator: a ring of eight small adjustable electromagnets round the beam.
A rule of thumb for a lens whose field spreads over a few millimetres:
focal length ≈ 37 × V × S ÷ (NI)²
with V in volts, S the gap between pole pieces, NI the ampere-turns, and focal length in the same units as S. A lens with a 4 mm gap, 1,000 ampere-turns and 50 kV electrons has a focal length of about 7 mm. Push the current up and it shortens, but not below about half the gap: there the iron saturates and the rule fails.
Two kinds of microscope
The transmission electron microscope (TEM) works like a slide projector, upside down. Electrons from a gun at the top pass through a specimen so thin they mostly go straight through. Where it is thicker or made of heavier atoms, more electrons are scattered aside and stopped by the objective aperture, so that place is dark in the image. Lenses magnify the image in two or three stages onto a fluorescent screen.
The scanning electron microscope (SEM) works like a Cathode Rays and Television camera. Lenses squeeze the beam to a tiny spot on the surface of a solid object, and coils sweep the spot across it line by line. Where the spot lands, the surface throws off slow secondary electrons: more from edges and slopes, fewer from hollows. A detector counts them and sets the brightness of a spot on a picture tube swept in step. The result looks like a photograph of a tiny landscape, with enormous depth of focus.
What you need
| Item | Use | Source |
|---|---|---|
| Diffusion-pumped vacuum, 10⁻⁴–10⁻⁵ mm, with O-ring flanges | The whole column | High-Vacuum Pumps and Gauges |
| Smooth, steady high-voltage supply, 30–60 kV | Electron gun | Power Supplies and Regulators, Particle Accelerators (Overview) (voltage multiplier) |
| Soft iron, brass, non-magnetic steel; lathe work to 0.01 mm, pole pieces to 0.002 mm | Lenses and column | The Lathe, Precision Grinding |
| Enamelled copper wire, several kilograms | Lens coils | Copper Wire |
| Tungsten wire 0.1–0.15 mm | Filaments | Tungsten and Filament Wire |
| Platinum foil | Apertures, shadowing | Platinum and the Platinum Metals |
| Zinc sulfide or zinc silicate phosphor | Viewing screen | Cathode Rays and Television |
| Fine-grained photographic plates | Recording images | Photography |
| Lead glass, sheet lead | Window and X-ray shield | Lead, Optical Glass and Achromatic Lenses |
| Stable current supplies, 0.2–1 A | Lenses | Power Supplies and Regulators |
| Oscilloscope parts, time bases, amplifiers | Scanning microscope | Cathode Rays and Television, Amplifiers and Oscillators |
Where to find it. Everything here is made, not gathered. The two raw materials worth naming: graphite (or pure charcoal) for carbon support films and coating, and platinum for apertures and shadowing. See Rocks and Minerals and Platinum and the Platinum Metals.
The parts
The electron gun
The same as the gun of a cathode-ray tube (Cathode Rays and Television), run at far higher voltage.
- Filament. Tungsten wire 0.12 mm thick, bent into a sharp V hairpin with a 0.1 mm radius at the tip, spot-welded to two stiff legs set in glass or ceramic. Run it at about 2.5 A, white-hot (about 2,700 °C). It lasts 30–100 hours; keep a dozen spares.
- Grid (Wehnelt cap). A metal cup round the filament with a 1–2 mm hole just in front of the tip, held 100–500 V more negative than the filament. A resistor of a few million ohms between them makes the bias set itself. It squeezes the electrons into a small crossover, the real source the lenses image.
- Anode. An earthed plate with a 3–5 mm hole, 10–15 mm below the cap.
- High voltage. The filament and cap sit at −30 to −60 kV; everything else, including you, is at earth. The filament's heating current comes from a small transformer insulated for 80 kV, sunk in oil with the supply (The Transformer).
A tungsten hairpin needs only 10⁻⁴ mm. A sharp single crystal of tungsten used as a field emitter gives a thousand times brighter beam but needs 10⁻⁹ mm and ion pumps (Ion Pumps and Leak Detectors).
The magnetic lens
- Case. Turn a pot of soft, annealed iron, about 12 cm across and 6 cm tall, with a bore through its centre and a removable lid (The Lathe).
- Pole pieces. Two short soft-iron cones (iron with 50 % cobalt is better), the bore through them 3–5 mm, the gap between their tips 3–5 mm, kept apart by a brass spacer. They must be truly round and on one axis to two thousandths of a millimetre; grind and lap them on the grinder in one setting. Any error shows up as astigmatism.
- Coil. 2,000–4,000 turns of 0.5–0.6 mm enamelled copper wire on a brass bobbin, carrying up to 1 A. It gives off tens of watts: wrap a copper water-cooling coil round it, or the drift as it warms will defocus the image.
- Supply. A battery, or a regulated supply steady to 1 part in 10,000 or better (Power Supplies and Regulators). A coarse and a fine control in series.
- Centring. Mount each lens so it can be slid sideways by two screws at right angles, through O-ring seals (Packing, Gaskets and Seals).
The column
A vertical stack, 1–1.5 m tall, of soft-iron and brass sections joined by O-ring flanges, so it can be taken apart. The iron also screens stray magnetic fields. From the top:
| Part | What it does |
|---|---|
| Gun with tilt and shift screws | Makes the beam |
| Condenser lens (weak) | Gathers the beam onto a spot 10–50 µm wide on the specimen |
| Specimen holder | A rod carrying the 3 mm grid into the objective gap through a small airlock |
| Objective lens and aperture | Forms the first image, ×50–100 |
| Intermediate lens (later) | Changes the total magnification |
| Projector lens | Enlarges again, ×50–200 |
| Viewing chamber | Fluorescent screen behind a lead-glass window, a ×10 magnifier for focusing |
| Camera | Plates under the screen, which tips up; its own airlock |
| Pumping port | Large and short, low on the column |
Total magnification is the product of the stages: ×80 from the objective and ×100 from the projector make ×8,000. Measure it rather than trust the arithmetic (below).
Where to put it. On a massive concrete block on the ground floor, away from lorries, machinery, lifts and above all transformers and heavy wiring: the 50 Hz field round a mains cable a few metres away can shift the beam by more than the detail you want to see.
Aligning it
- Light the filament slowly; watch the screen for the first glow.
- Shift and tilt the gun until the brightest beam runs down the axis.
- Centre each lens in turn, from the top down: rock its current up and down (or rock the high voltage), and slide the lens until the image swells and shrinks about its centre instead of sliding sideways. This is the most important adjustment.
- Centre the objective aperture on the beam.
- Correct astigmatism. Look at a small round hole in a carbon film just out of focus: a thin bright fringe goes round its edge. Turn the stigmator until the fringe is even all round.
Making specimens
The beam is stopped by anything thicker than about 0.1 µm at 50 kV. Everything is dried and goes into the vacuum.
Grids and support films
- Grids. Discs 3 mm across of fine copper mesh, 200–400 wires to the inch (about 60–120 µm squares). Punch them from fine woven bronze or copper gauze (Iron and Brass Wire), or electroplate copper onto a ruled pattern and peel it off (Electroplating).
- Collodion film. Let one drop of a 2 % solution of collodion (nitrocellulose, Early Plastics) in amyl acetate fall onto clean water in a dish. It spreads into a film 20–30 nm thick, showing faint grey. Lay the grids on it, then lift the lot onto a filter paper and dry.
- Carbon film. Stronger, and stands the beam better. In a vacuum coater, make a small arc between two pointed graphite rods pressed together; carbon flies off and settles on a sheet of freshly split mica. Float the film off on water and pick it up on grids.
Particles and viruses
Suspend them in distilled water, put a drop on a filmed grid, wait a minute, wick it off with filter paper and let it dry. Small, light particles show little contrast. Two old tricks:
- Negative stain. Mix the drop with a 1–2 % solution of a heavy-metal salt — phosphotungstic acid (Tungsten and Filament Wire) or uranyl acetate. It dries into a dark glass round each particle, which shows light on dark with its surface detail picked out. This is how most viruses were first seen and classified.
- Shadowing. In the vacuum coater, evaporate a few milligrams of platinum or chromium from a tungsten wire held low to one side, so it strikes the grid at 10–20°. Each particle casts a "shadow" with no metal in it. Shadow length × the tangent of the angle gives its height. Print the image as a negative and it looks lit from one side.
Surfaces: replicas
A solid lump of metal, a fracture or a tooth cannot go in a TEM. Coat its surface with collodion or evaporated carbon, shadow it, then dissolve or strip the original away in acid. The thin replica, a cast of the surface, goes on a grid.
Thin sections of tissue
The extension of Tissue Slides and the Microtome to slices a hundred times thinner.
- Fix 1 mm cubes of fresh tissue in buffered formaldehyde or glutaraldehyde (Methanol and Formaldehyde), then in 1 % osmium tetroxide, which fixes fats and stains them black. Osmium tetroxide is very dangerous (see Dangers).
- Dehydrate through 30, 50, 70, 90 and 100 % alcohol, 10 minutes each.
- Embed in epoxy or methacrylate resin, cured hard in an oven at about 60 °C for a day (Resins: Phenolic, Polyester and Epoxy).
- Glass knife. Score a strip of plate glass 6 mm thick and 25 mm wide into squares, then score a square across the diagonal and break it with even pressure. The fresh edge, if the break is clean, is sharp to a few nanometres. Look at it under the light microscope: use only the straight part of the edge, free of nicks. Fix a small trough of tape behind the edge and fill it with water.
- Ultramicrotome. The resin block is carried on a stiff arm that swings down past the knife. The advance per cut — 50–100 nm — comes not from a screw but from the slow heat expansion of a metal rod warmed by a small lamp. Sections float off onto the water.
- Judge thickness by colour. Reflected light on the floating sections shows interference colours: grey under 60 nm, silver 60–90 nm, gold 90–150 nm. Keep the silver ones.
- Pick them up on filmed grids; stain with uranyl acetate and lead citrate for contrast.
Metals and minerals
Punch a 3 mm disc from foil, and thin its centre by electropolishing in an acid bath until a tiny hole appears (Electroplating). The rim of the hole is thin enough to see crystal flaws, grain boundaries and fine precipitates.
The scanning electron microscope
The SEM needs no thin specimens and no magnifying stages, and much of its electronics is that of Cathode Rays and Television. Its hard parts are the very fine spot and the sensitive detector.
- Gun as above at 5–30 kV.
- Lenses. Two or three lenses demagnify the gun's crossover, about 50 µm wide, a thousand times or more. The last lens has a long focal length, 10–20 mm, so there is room below it for the sample.
- Scan coils. Two pairs of small coils inside the bore of the last lens deflect the beam in X and Y, fed by sawtooth currents from two time bases: a fast line sweep and a slow frame sweep. For looking, 50–100 lines a second; for a photograph, 1,000 lines in 1–2 minutes.
- Display. The same sawtooth voltages sweep a picture tube. Magnification = width of the picture ÷ width scanned. A 10 cm picture of a 10 µm square is ×10,000. Turning down the scan currents with a switch of resistors changes the magnification without touching the lenses.
- Detector. A metal cage at +250 V beside the sample pulls in the slow secondary electrons. Inside it, a small disc of plastic or a zinc sulfide scintillator, coated with a film of aluminium and held at +10 kV, turns each electron into a flash of light. A rod of clear glass or acrylic carries the light out of the vacuum to a photomultiplier (Photomultipliers and Scintillation Counters): a tube whose light-sensitive cathode gives off electrons that are multiplied a million times by a chain of electrodes, each at about 100 V more than the last. A plain collector plate and electrometer (Amplifiers and Oscillators) work, slowly.
- Picture. The amplified detector signal drives the grid of the picture tube. Photograph the screen during a slow scan.
Samples must conduct, or charge builds up and the picture flares. Coat them with 10–20 nm of gold in a sputter coater or with evaporated carbon. Living material must first be fixed, dehydrated and dried without the surface tension of evaporating water crushing it — freeze-drying (Preserving Food) is the simplest way.
Depth of focus. The beam converges at only about a hundredth of a radian, so a picture at 1 µm resolution is sharp through a depth of about 100 µm. Insects, pollen, fractured metal and broken teeth look solid, which no light microscope can do.
Stepping stones
- The electron-optical bench. A demountable cathode-ray tube on a diffusion pump (Cathode Rays and Television), with a fine wire mesh across the beam and a fluorescent screen 30 cm below. One movable focusing coil makes a sharp, enlarged, rotated image of the mesh. Add a second coil and the image is magnified twice over. This was the first electron microscope (×14, in 1931). It teaches alignment, stray fields and vacuum practice.
- The two-lens TEM at 30–50 kV: objective and projector, a specimen holder and a screen. ×2,000–10,000. It beats the light microscope once it resolves better than 0.2 µm: test it on shadowed particles and diffraction-grating replicas.
- The three-lens TEM with condenser, aperture, stigmator, airlocks and a camera. Viruses, bacteria in section, metal foils. ×10,000–50,000 at 10–20 nm.
- The SEM. With magnetic lenses and time bases already mastered, a scanning instrument at 1 µm resolution is easier than a TEM and already shows surfaces no light microscope can. Improve the spot and the detector from there.
Electron diffraction. Weaken the projector so it shows the objective's back focal plane instead of the image, and a crystal gives a pattern of spots or rings whose spacing gives the spacing of its atoms — a way to identify minerals and metal phases.
How to tell it worked
- Calibrate magnification with a replica of a ruled diffraction grating of known spacing — a grating of 2,160 lines per millimetre has lines 463 nm apart — or with plastic spheres of known size.
- Resolution. Find the narrowest gap you can clearly see between two shadowed particles, or the finest detail on a known specimen. Tobacco mosaic virus, from the mottled leaves of infected tobacco or tomato plants, is a rod 18 nm wide and 300 nm long: seeing its width means you have resolved better than 20 nm.
- Stability. A photograph exposed for 4 seconds is as sharp as one exposed for 1.
Note Troubleshooting. No beam at all → filament burnt out, gun far off axis, or vacuum too poor (the high voltage arcs and trips). Image slides sideways when you focus → lens off centre; repeat the current-rocking alignment. Image blurred one way more than the other → astigmatism; use the stigmator; clean the aperture, which charges up when dirty. Image wobbles or is doubled → 50 Hz stray field or vibration; move cables and transformers, shield with soft iron. Image drifts out of focus → lens coils warming, or a high-voltage supply that wanders; cool and regulate. Specimen film bursts or particles move → beam too intense; spread it with the condenser; use carbon film. SEM picture flares and streaks → sample charging; coat it.
Dangers
Danger The gun runs at 30–100 kV, which kills. The supply and its oil tank must be fully enclosed and earthed, with an interlock that switches off and earths the output when any cover is opened. Earth the filament end with an earthing rod before touching it, every time.
Danger Electron microscopes make X-rays wherever the beam strikes an aperture, the specimen or the screen (X-Rays). Use lead glass for every window, line the viewing chamber with lead, and check the column with photographic film or a Geiger counter for leaks before routine use (Detecting Radiation).
Danger Osmium tetroxide vapour fixes the surface of the eye as it fixes tissue: it can blind, and it damages the lungs. Keep it in sealed glass ampoules, open and use it only in a fume cupboard, and neutralise spills with corn oil or milk powder, which turn it black and harmless.
- Uranyl acetate is a poisonous heavy metal and slightly radioactive; lead citrate is poisonous. Gloves, no dust, careful waste.
- Epoxy resins and their hardeners cause allergic skin rashes after repeated contact; gloves.
- Glass knives and broken glass cut deeply.
- Liquid nitrogen in traps, implosion of windows and mercury pumps: as in High-Vacuum Pumps and Gauges.
What this unlocks
Seeing viruses and the structure of cells (Infectious Diseases, Vaccination, Heredity and Genetics, Cells, Life and Evolution); judging metals, ceramics, cements and catalysts by their microstructure (Strength of Materials, Portland Cement); checking and repairing the patterns on chips (Making Microchips). The SEM's fine, steerable beam leads directly to electron-beam lithography, which writes chip patterns finer than light can, and to electron-beam welding and melting (Welding).
How it was first done
Louis de Broglie proposed in 1924 that electrons have a wavelength, and electron diffraction proved it in 1927. Hans Busch showed in 1926 that a short coil focuses electrons like a lens. At the Technical University of Berlin, Max Knoll and his student Ernst Ruska built a two-lens instrument in 1931 that magnified a wire mesh 14 times. By 1933 Ruska, with iron-clad pole-piece lenses, had passed the resolution of the light microscope. Ruska and Bodo von Borries built the first production instruments at Siemens in 1939, and in the same year Ruska's brother Helmut, with Gustav Kausche and Edgar Pfankuch, published the first pictures of tobacco mosaic virus. In Toronto, Albert Prebus and James Hillier built one in 1938; Hillier went on to design RCA's instruments. Manfred von Ardenne built a scanning microscope in Berlin in 1937–38. Charles Oatley's group at Cambridge developed it from 1948, Thomas Everhart and Richard Thornley devised the scintillator detector in 1960, and the first commercial SEM, the Cambridge Stereoscan, came out in 1965. Ruska shared the Nobel Prize in Physics in 1986, fifty-five years after his first instrument.