Photomultipliers and Scintillation Counters

Some materials give off a tiny flash of light when a particle of radiation passes through them. Count the flashes and you count the radiation; measure how bright each one is and you learn the energy, and so which element made it. The spinthariscope you can build this week, the crystals that flash, the photomultiplier tube that turns one flash into a measurable pulse, and the counter that does the work of prospecting, medicine and physics.

Difficulty
●●●●●
Time
An evening for a spinthariscope; weeks for a plastic scintillator; a year or more of vacuum work for a photomultiplier tube
Steps from zero
49 prerequisites →

You need first

Detecting RadiationThe Vacuum TubePhotocells and Sensors

This unlocks

—
The full path from nothing — 49 steps before this one
  1. Fire
  2. Cordage and Rope
  3. Stone Tools
  4. Finding and Preparing Clay
  5. Finding and Identifying Ores
  6. Sulfur
  7. Hunting and Trapping
  8. Basic Woodworking
  9. The Kiln
  10. Pottery
  11. Charcoal
  12. Hides and Leather
  13. Lime
  14. Bricks
  15. Salt
  16. Mercury
  17. Lye and Potash
  18. Fermentation
  19. The Wheel and Axle
  20. Glass
  21. Firebrick and Refractories
  22. Bellows
  23. Saltpeter
  24. Acids, Bases and Testing
  25. The Lathe
  26. Gears
  27. Copper Smelting
  28. Iron (the Bloomery)
  29. Laboratory Glassware
  30. Sulfuric Acid
  31. Pumps
  32. The Waterwheel
  33. Static Electricity and the Leyden Jar
  34. Blacksmithing
  35. Hydrochloric Acid
  36. Identifying Substances
  37. The Battery
  38. The Vacuum Pump
  39. Magnets
  40. Selenium and Tellurium
  41. The Light Bulb
  42. Copper Wire
  43. The Generator
  44. The Vacuum Tube
  45. The Electric Arc Furnace
  46. Detecting Radiation
  47. Pure Silicon
  48. Semiconductors and the Transistor
  49. Photocells and Sensors
  50. Photomultipliers and Scintillation Counters

Do them roughly in this order. See it on the tech tree →

Why it matters

The Geiger counter of Detecting Radiation answers one question: is there radiation here? A scintillation counter answers three more:

  • How much? A crystal of sodium iodide the size of a plum stops most of the gamma rays that pass through it, where a Geiger tube stops perhaps one in a hundred. For gamma rays, a scintillation counter is ten to fifty times more sensitive, which turns an afternoon of searching into a minute of it.
  • What energy? Each flash is as bright as the energy the radiation deposited. Sort the flashes by brightness and you get a spectrum with peaks at particular energies — and because every radioactive element emits gamma rays at its own energies, the peaks tell you which element: uranium, thorium or potassium.
  • How fast? A plastic scintillator responds in a few billionths of a second, hundreds of times faster than a Geiger tube's dead time, so it can count intense radiation without choking and can time a particle's flight.

That combination makes the scintillation counter the workhorse detector of the modern world: the prospector's scintillometer and borehole logger (Mining and Tunnelling, Finding and Identifying Ores, Petroleum and Refining); the thyroid probe and gamma camera of Ultrasound, CT and MRI; industrial thickness and level gauges; the liquid scintillation counter that measures carbon-14 (Cells, Life and Evolution, Heredity and Genetics); and the portal monitors that catch a lost radiotherapy source on its way into a scrap furnace (Radiotherapy (Overview)).

How it works

From one gamma ray to one measurable pulse gamma ray crystal: the ray stops here and 25,000 photons flash out in a quarter of a millionth of a second photomultiplier: about one photon in four frees an electron from the photocathode, and each dynode turns one electron into three to five. Ten stages, a gain of a million. height of each pulse ∝ the energy the particle left in the crystal. Count the pulses and you have a counter; sort them by height and you have a spectrometer. lightanodeload
Fig. 1 The chain from particle to pulse. Radiation deposits energy in the crystal; a fraction of it comes out as a flash of visible light; the light reaches a photocathode and knocks out a handful of electrons; each electron is multiplied at a series of dynodes until a million arrive at the anode as a measurable pulse. The pulse's height is proportional to the energy the particle deposited.

1. The flash. Radiation passing through certain transparent materials excites their atoms or molecules. Most of the energy becomes heat, but in a good scintillator a few percent comes back out as visible or near-ultraviolet light. The numbers are small but workable: sodium iodide with a trace of thallium gives about 38 photons per keV of energy deposited, so a 662 keV gamma ray that stops completely in the crystal produces about 25,000 photons in a flash lasting a quarter of a microsecond.

2. Getting the light out. Those photons must reach the detector, so the crystal must be transparent to its own light, wrapped in a white reflector so stray photons bounce back, and optically coupled to the photo-detector with a grease or oil of similar refractive index. An air gap of a thousandth of a millimetre loses half the light to reflection.

3. Turning light into electrons. Light falling on certain surfaces — a film of caesium and antimony is the classic — knocks electrons out into a vacuum. This is the photoelectric effect, and the surface is a photocathode (Photocells and Sensors). The best photocathodes convert only about one photon in four to five into a useful electron, so our 25,000 photons become perhaps 5,000 electrons — still far too few to measure.

4. Multiplying them. An electron accelerated through 100 volts and slammed into the right surface knocks out three to five more. Do that ten times in series and one electron becomes three to five million. Each stage is a dynode, and the chain of them is what makes a photomultiplier tube. The tube's output is a pulse of current large enough for an ordinary valve or transistor amplifier.

5. Measuring the pulse. Count the pulses and you have a counter. Sort them by height and you have a spectrometer, because pulse height is proportional to the energy deposited in the crystal.

What you need

Thing For Chapter
Zinc sulfide, silver nitrate, a crucible and a 1,000 °C furnace The simplest scintillator Zinc and Brass, The Kiln, Very-High-Temperature Furnaces
Iodine from kelp, sodium carbonate Sodium iodide crystals Iodine, Soda Ash and Baking Soda
A sealed furnace, dry inert gas, a slow lowering mechanism Growing a crystal Very-High-Temperature Furnaces, Liquid Air, Nitrogen and the Noble Gases
Styrene, or naphthalene and anthracene from coal tar Plastic and organic scintillators Petrochemicals and Modern Plastics, Coal Tar Chemicals
Full glassblowing and high-vacuum practice The photomultiplier Scientific Glassblowing and Glass-to-Metal Seals, The Vacuum Tube, High-Vacuum Pumps and Gauges, Ion Pumps and Leak Detectors
Caesium and antimony The photocathode and dynodes Sodium and Potassium Metal, Antimony and Bismuth
A 1,000–1,500 V stable supply, 10–14 resistors, capacitors Running the tube Power Supplies and Regulators, Resistors, Capacitors and Coils
Amplifier, discriminator, scaler, timer Counting Amplifiers and Oscillators, Electronic Counters
Mu-metal or thick soft iron tube Magnetic shield Transformer Steel, Ferrite and Magnetic Shielding, Permanent Magnets

Method 1: The spinthariscope — see single atoms decay

What it is

A spinthariscope is the simplest radiation detector there is: a little tube with a zinc sulfide screen at one end, a tiny radioactive source just in front of it, and a magnifier to look through. In the dark, after your eyes have adapted, you see individual flashes — each one a single atomic nucleus disintegrating and firing an alpha particle into the screen.

The spinthariscope, in section zinc sulfide screen (silver-activated), one or two grains deep a speck of uranium or thorium mineral on a needle point, 2–3 mm from the screen. Alpha particles travel only a few centimetres in air, so it must be this close. blackened tube of card or wood lens, 15–40 mm focal length the dark-adapted eye: sit in complete darkness for 20 minutes first Each brief blue-white spark is one atomic nucleus disintegrating. Ten to twenty a minute is the usual sight. Rutherford'sgroup counted hundreds of thousands of these by eye, and found the atomic nucleus by doing it.
Fig. 2 The spinthariscope. A blackened tube holds a zinc sulfide screen at the far end, a speck of alpha-emitting mineral on a needle a few millimetres in front of it, and a simple lens at the eye end. In full darkness, after twenty minutes of dark adaptation, each flash is one atom disintegrating.

Making it

  1. The screen. Make the phosphor: precipitate zinc sulfide by passing hydrogen sulfide through a solution of a zinc salt (Zinc and Brass, Sulfur), wash it very thoroughly and dry it. Mix in about 1 part in 10,000 of silver nitrate dissolved in a little water, plus about 2% common salt as a flux, dry the paste and fire it in a covered crucible at 950–1,000 °C for 30–60 minutes, with a little sulfur or charcoal in the crucible to keep the atmosphere from oxidising the sulfide (The Kiln, Very-High-Temperature Furnaces, Glazes for crucible practice). Cool, break up the cake gently, and sieve. The powder should be pale cream and glow brightly green-blue under an ultraviolet source or near a radioactive mineral.
  2. Coat a disc. Brush a very thin layer of the powder onto a glass or mica disc with a trace of thin gum or water glass as binder — one or two grains deep, no more. A thick screen hides its own flashes.
  3. The source. The point of a needle touched to a crushed uranium or thorium mineral (Detecting Radiation) and fixed 2–3 mm from the screen. Alpha particles travel only 3–7 cm in air and are stopped by paper, so the source must be very close and uncovered.
  4. The tube. A tube of blackened card or wood, with a lens of 15–40 mm focal length at the eye end (Lenses), focused on the screen.

Danger Use only a naturally radioactive mineral in a fixed, sealed mount, and never scrape, grind or open an old luminous dial. Old clock and instrument dials were painted with radium compounds, and that paint powders. Breathing or swallowing it puts an alpha emitter inside your bones for the rest of your life — the fate of the dial painters of the 1920s, who died of jaw and bone cancers (Watches and Small Clocks, Rare Earths, Cerium and Phosphors). A sealed mineral specimen on a needle, handled with tongs and kept in a tin outside the house, is safe; dust is not.

Dark adaptation is everything. Sit in complete darkness for 20 minutes before looking. A sprinkling of ten or twenty flashes a minute, each a brief blue-white spark, is the usual sight.

Method 2: The scintillator crystal

Sodium iodide with thallium

The standard gamma-ray scintillator is a single crystal of sodium iodide containing about 0.1% thallium iodide, written NaI(Tl). It is excellent and it is awkward: it dissolves in its own weight of water, it fogs in damp air in minutes, and thallium is poisonous.

It is worth the trouble: iodine's heavy atoms stop gamma rays well, and its bright, clear light gives a usable energy measurement — a good crystal separates two gamma energies 7 to 9% apart.

Growing one (the Stockbarger–Bridgman method, which is also how Fused Quartz and Halogen Lamps and Synthetic Gems crystals are grown):

  1. Dry the sodium iodide hard — hours at 150 °C under vacuum — and add the thallium iodide, 1 g per kilogram.
  2. Load it into a sealed crucible with a pointed bottom, inside a furnace with a hot zone above 661 °C (the melting point) and a cold zone below it, under dry inert gas or vacuum.
  3. Lower the crucible through the temperature gradient at 1–3 mm an hour. Freezing begins at the point and a single crystal grows upward from it. A 50 mm crystal takes several days.
  4. Anneal by cooling at 10–20 °C an hour; faster cooling cracks it.
  5. Cut and polish in a dry box, under mineral oil, with saws and laps wetted with oil, never water (Gem Cutting and Lapidary, Precision Grinding).
  6. Can it. Wrap in white reflector — magnesium oxide powder or several layers of fine white cloth — and seal in a thin aluminium can with a glass window at one end, the window cemented in with a moisture-proof resin. The can is permanent; once air reaches the crystal, it is finished.

A 38 mm diameter by 38 mm long crystal is the classic general-purpose size; a 25 mm one suffices for prospecting, a 75 mm one for weak laboratory sources.

The plastic scintillator

A plastic scintillator is far easier, needs no dry box, can be cast to any shape and size, and responds in nanoseconds. Its weakness is that it is made of light atoms, so it hardly stops gamma rays at all — it is a beta, alpha and fast-neutron detector, and a very good one.

Make it by polymerising styrene (Petrochemicals and Modern Plastics) with two dissolved dyes:

  • about 1–2% of an aromatic scintillating compound — p-terphenyl is the classic; anthracene or naphthalene from Coal Tar Chemicals will work less well;
  • 0.02–0.05% of a wavelength shifter, a second dye that absorbs the first one's ultraviolet and re-emits visible blue light that glass and photocathodes accept.

Dissolve the dyes in the purified, inhibitor-free monomer, filter, pour into a glass or polished metal mould, exclude air, and cure at 70 °C for two days, then 100–120 °C for a day. Cool slowly over a day to avoid internal strain, then machine and polish the faces like a lens (Lenses). A sheet 5–20 mm thick of any area, or a block, works. Keep it away from solvents and oils, which craze it.

A crystal of anthracene grown from the melt (mp 216 °C) is the traditional organic standard, about half as bright as NaI(Tl) and far easier to grow — a good stepping stone.

Which scintillator for what

Material Light output Flash length Best for
Zinc sulfide (silver) powder Very high ~200 ns Alpha only — opaque, so a thin layer
NaI(Tl) crystal 38 photons/keV 230 ns Gamma rays and energy measurement
Caesium iodide (thallium) 50–60 photons/keV 1,000 ns Gamma; tougher and less hygroscopic, slower
Anthracene crystal ~17 photons/keV 30 ns Beta particles; the organic reference
Plastic ~10 photons/keV 2–3 ns Beta, fast counting, large areas, odd shapes
Liquid (solvent + dyes) ~8 photons/keV 2–4 ns Weak beta emitters dissolved in the liquid — carbon-14, tritium

Method 3: The photomultiplier tube

Be honest about the difficulty

A photomultiplier tube is the hardest vacuum tube in this book. It needs everything in The Vacuum Tube plus a vacuum better than 10⁻⁶ of an atmosphere, a photocathode formed by distilling caesium onto antimony inside the finished, sealed envelope, a dozen electrodes positioned to a few tenths of a millimetre, and a complete absence of light leaks and leakage currents. Expect a long apprenticeship and many failures.

Three ways round it, in order of usefulness:

  1. Count flashes by eye — the spinthariscope, as Rutherford's group did. It is tedious and it works.
  2. Use a single-stage photocell instead, with a very high-gain amplifier (Photocells and Sensors, Amplifiers and Oscillators). You lose the ability to see single gamma rays but can measure a strong, steady glow — enough for a thickness gauge or a bright beta source.
  3. Use a silicon photodiode, once Semiconductors and the Transistor and Pure Silicon exist. A large-area photodiode on a NaI crystal, with a low-noise charge amplifier, makes a working spectrometer with no vacuum at all, and is how many modern portable instruments are built. This is the route a rebuilding society should plan for.

What is inside one

Inside a photomultiplier tube, and the chain that feeds it end window and the photocathode of Sb-Cs dynodes, 10–12 of them, each about 100 V more positive than the last anode to the amplifier focusing electrode −1,000 to −1,500 V Equal resistors of 100 kΩ to 1 MΩ divide the supply between the stages. The chain must draw at least 100 times the peak signal current, or the gain sags on big pulses; the capacitors on the last three stages supply the instant current of each pulse. to anode load
Fig. 3 Inside a photomultiplier tube, and the resistor chain that runs it. Light enters through the end window onto the semi-transparent photocathode. A focusing electrode steers the photoelectrons onto the first dynode, and each dynode is held about 100 V more positive than the last by the divider chain, so electrons cascade down the line. Capacitors on the last few stages supply the current for each pulse.
  1. The envelope: a glass tube, typically 40–55 mm across and 100–150 mm long, with a flat end window; ultraviolet scintillators need a quartz window (Fused Quartz and Halogen Lamps).
  2. The photocathode: a semi-transparent film of antimony with caesium (Sb-Cs), deposited on the inside of the window by evaporating antimony from a small filament inside the tube and then admitting caesium vapour from a side capsule, with the tube warm, until the photoelectric current peaks. It converts about 20–25% of blue photons into electrons.
  3. A focusing electrode at a few hundred volts, shaping the field so that photoelectrons from every part of the window land on the first dynode.
  4. Ten to twelve dynodes, arranged as a curved "venetian blind" or a staggered "box and grid" so that electrons cannot fly past a stage. Each is a nickel or copper-beryllium plate whose surface is activated — in a caesium tube, the same caesium-antimony treatment; in a tougher tube, beryllium-copper oxidised to grow a beryllium oxide film. Each gives 3 to 5 secondary electrons per arriving electron.
  5. The anode, a mesh or plate in front of the last dynode, from which the pulse is taken.
  6. A getter fired after sealing, as in any valve, to clean up the last gas (The Vacuum Tube).

Running it

  • Voltage: a stable, well-smoothed 1,000–1,500 V supply (Power Supplies and Regulators). The gain changes by about 10% for every 1% change in supply voltage, so the supply must be steadier than any other in the laboratory — better than 0.1%.
  • The divider chain: equal resistors of 100 kΩ to 1 MΩ between successive dynodes, drawing a bleeder current at least 100 times the peak signal current. Put 10–50 nF capacitors across the last three or four stages to supply the instantaneous current of each pulse; without them the gain sags on big pulses.
  • Polarity: either run the cathode at a large negative voltage and the anode near earth (convenient for the signal, dangerous at the window), or the cathode at earth and the anode at +1,200 V (safer, but the output must be taken through a capacitor). Never let the glass near the photocathode sit at a different potential from the cathode film — the leakage current through the glass produces light and noise.
  • Darkness: the tube must be in a light-tight housing, and never, ever switched on in room light. A photomultiplier exposed to daylight with volts on it is destroyed in seconds. Wire a lid interlock if you can.
  • Magnetic shielding: a few hundredths of a tesla — the field near a loudspeaker or a transformer — deflects the electrons enough to cut the gain severely. Surround the tube with a tube of mu-metal or annealed soft iron, extending past the photocathode, connected to the cathode potential (Transformer Steel, Ferrite and Magnetic Shielding).

The scintillation counter and the pulse height analyser

Assembling a counter

A scintillation counter is the crystal, the tube and the electronics in one light-tight, magnetically shielded housing:

  1. Crystal can against the photomultiplier window, with a thin layer of optical grease — clear silicone grease (Silicones and Synthetic Rubbers) or a drop of mineral oil — squeezed out to exclude all air.
  2. Spring-loaded mount so the crystal stays pressed to the window as things expand, with soft rubber round the joint and black tape over everything.
  3. Mu-metal shield, then an aluminium or brass housing, light-tight at every joint, with a thin aluminium or beryllium entrance window if you want to detect low-energy radiation.
  4. A cathode follower or emitter follower mounted inside the housing, within a few centimetres of the anode, to drive the cable without picking up hum (Amplifiers and Oscillators, Interference, Screening and Earthing).
  5. A linear amplifier of adjustable gain, which must be linear — if it distorts, the energy information is destroyed.
  6. A discriminator (a Schmitt trigger: see Electronic Counters) that produces one standard pulse for every input pulse above a set height, and a scaler and timer to count them for a measured time.

The pulse height analyser

A pulse height analyser sorts pulses by height, and in its simplest useful form — the single-channel analyser — it is two discriminators and one logic gate:

  • Discriminator A fires on pulses above a level you set, the threshold.
  • Discriminator B fires on pulses above threshold + window width.
  • A gate passes a count only when A fired and B did not — that is, only for pulses inside the window.

Build it from two Schmitt triggers and an anticoincidence gate (Electronic Counters, Logic and Binary). Now step the threshold up in equal steps, counting for a fixed time at each setting, and plot counts against threshold. The plot is a gamma-ray spectrum, with a bump — the photopeak — at each gamma energy the source emits. Measuring it by hand takes an hour per spectrum; the multichannel analyser that does it all at once needs a fast analogue-to-digital converter and memory (Microprocessors and Personal Computers, The Computer).

Calibrating the energy scale with no standards to buy. Nature supplies reference lines:

Source Energy Where to get it
Potassium-40 1,461 keV A bag of any potash fertiliser or potassium salt (potash fertiliser) — weak but perfectly steady
Thallium-208 (thorium series) 2,615 keV Thorium minerals, old gas mantles (Rare Earths, Cerium and Phosphors)
Radium-226 and daughters 186, 352, 609 keV Uranium minerals
Caesium-137 662 keV A standard, if the old world left you a check source

Plot peak position against known energy: it should be a straight line through the origin. If it is not, your amplifier is non-linear.

How to tell it worked

  • The dark count. With no source and the lid shut, a good counter gives a few tens to a few hundred counts a minute at a threshold just above noise, falling steeply as you raise the threshold. A rattle of thousands means a light leak, a cracked tube or too much voltage. Shut the tube in the dark for an hour before judging: a tube just exposed to room light sulks for hours.
  • Setting the voltage. There is no Geiger plateau to look for. Set the amplifier gain and threshold, put a source in place, and raise the voltage until a photopeak sits where you want it on the scale.
  • Energy resolution. Measure a photopeak's width at half its height and divide by the energy at its centre. A good NaI(Tl) counter gives 7–9% at 662 keV; 12% is poor but usable; 25% means the crystal is fogged, the optical coupling has an air gap, or the tube's photocathode is poor.
  • Linearity. Two sources of known, different energies must give peaks whose positions are in the same ratio as their energies.
  • Sensitivity in the field. A 25 mm NaI counter waved over ordinary granite should read noticeably above its reading over a river's water or a wooden floor. If it does not, your gain is too low.

Note Troubleshooting. Flashes visible in the spinthariscope only at the edge of vision → normal; the eye's rods are off-centre. Look slightly to one side of the screen. No flashes at all → screen too thick, source too far away (alpha range is centimetres), screen poorly fired, or not enough dark adaptation. Counter works for minutes then rate collapses → gain sag from a missing divider capacitor, or the supply current is too small for the bleeder chain. Photopeak drifts up or down over an hour → temperature. NaI(Tl) light output changes about 0.3% per °C and photomultiplier gain even more. Lag the housing; recalibrate on a check source at each session. Spectrum has peaks where no source should produce them → sum peaks (two gammas arriving together), the escape peak of iodine's own X-ray 28 keV below the photopeak, or mains interference being counted as pulses. Shield and earth properly (Interference, Screening and Earthing). Crystal has gone milky or yellow at the edges → water has reached it. The can has failed. It cannot be rescued; melt it down and regrow. Resolution fine at first, worse after re-mounting → air in the optical joint. Remake with more grease and no bubbles.

Dangers

Danger 1,000 volts at the back of the tube. It is a smoothed supply with capacitors across it that stay charged after switch-off, and it will kill. Earth the housing, enclose the divider chain, discharge the supply through a resistor and then short it before touching anything, and keep one hand in your pocket while probing (Wiring and Electrical Safety, Resistors, Capacitors and Coils).

  • Thallium compounds are extremely poisonous and are absorbed through skin as well as swallowed. Weigh them in a fume draught, wear gloves, never in a kitchen, and label every trace. If you cannot handle thallium safely, grow a caesium iodide or anthracene crystal instead.
  • Radioactive sources. Everything in Detecting Radiation's danger box applies: small, few, sealed, labelled, outside the house, handled with tongs, never ground or heated. Never open an unknown sealed source — see the warning in Radiotherapy (Overview).
  • Hydrogen sulfide, used to precipitate zinc sulfide, is as poisonous as hydrogen cyanide and deadens your sense of smell as it builds up. Outdoors or in a strong draught only (Sulfur).
  • Styrene and its vapour are flammable and harmful to breathe; polymerisation generates heat and can run away in a large casting. Cast in thin sections, keep the bath temperature controlled, and work in a draught (Petrochemicals and Modern Plastics, Ether, Acetone and Solvents).
  • A photomultiplier implodes if cracked while evacuated, throwing glass. Handle it in a cloth and wear eye protection (Scientific Glassblowing and Glass-to-Metal Seals).

What this unlocks

Gamma spectrometry is a chemical analysis that needs no chemistry: point a counter at a rock and read off its uranium, thorium and potassium. That makes serious prospecting possible (Finding and Identifying Ores, Mining and Tunnelling), and in a borehole it logs the strata as the probe goes down (Drills and Boring, Petroleum and Refining).

In medicine the same counter over a patient's neck measures how much iodine their thyroid takes up, and scanning it across the body builds the first radioisotope pictures; an array of them behind a lead plate full of holes is the gamma camera (Ultrasound, CT and MRI). In industry it gauges thickness, density and level through steel walls. In physics, two fast plastic counters in coincidence — counting only events seen by both within nanoseconds — pick a single kind of event out of a storm of background, which is the basic trick of all particle physics (Particle Accelerators (Overview)).

How it was first done

Scintillation counting is older than electronics. William Crookes built the first spinthariscope in 1903, and from 1908 Hans Geiger and Ernest Marsden, working for Ernest Rutherford in Manchester, counted alpha particles by eye as flashes on zinc sulfide screens, in shifts, in a darkened room. Those counts, showing that about one alpha in eight thousand bounced back off a gold foil, revealed the atomic nucleus in 1911.

The photomultiplier came from television research. Secondary-emission multiplication was patented by Joseph Slepian in 1919; Leonid Kubetsky in Leningrad built multi-stage multiplier tubes in 1930–34; and Vladimir Zworykin and colleagues at RCA produced practical photomultipliers from 1935–36. During the Second World War Samuel Curran and W. R. Baker in Berkeley put a photomultiplier behind a scintillating screen and counted the pulses electronically — the modern scintillation counter — though their work stayed secret until 1948.

Hartmut Kallmann in Berlin showed in 1947 that naphthalene scintillated well enough to count gamma rays, and Robert Hofstadter at Princeton reported in 1948 that sodium iodide activated with thallium was far better, with a light output proportional to energy — the discovery that created gamma spectrometry. Within a few years prospectors were walking the Colorado Plateau with scintillometers, and in 1958 Hal Anger in Berkeley built the first gamma camera from one large NaI crystal and a bank of photomultipliers.

Source: content/entries/electronics/scintillation-counters.md